Circulation · 2012
Long-term clinical and echocardiographic follow-up of the Freestyle stentless aortic bioprosthesis.
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mohammadi Siamak, Tchana-Sato Vincent, Kalavrouziotis Dimitri, Voisine Pierre, Doyle Daniel, Baillot Richard, Sponga Sandro, Metras Jacques, Perron Jean, Dagenais François
Аннотация
BACKGROUND: Stentless aortic bioprostheses were designed to provide enhanced hemodynamic performance and potentially greater longevity. The present report describes the outcomes of patients with the Freestyle stentless bioprosthesis followed for ≤18 years.
METHODS AND RESULTS: Between 1993 and 2011, 430 patients underwent primary aortic valve replacement with a Freestyle bioprosthesis in the subcoronary position. Mean age was 68.2 ± 8.2 years. All of the clinical and echocardiographic data were collected prospectively. Mean overall follow-up was 9.1 ± 4.4 years and was complete in all of the patients. In-hospital mortality was 3.5% (n=15). Overall, 10- and 15-year survival were 60.7% and 35.0%, respectively. Fifty-one patients required reoperation during follow-up, including 27 for structural valve deterioration (SVD). Overall, freedom from reoperation was 91.0% and 75.0% at 10 and 15 years, whereas freedom from reoperation for SVD was 95.9% and 82.3%, respectively. At 10 and 15 years, freedom from reoperation for SVD was 94.0% and 62.6% for patients <60 years of age and 96.3% and 88.4% for patients ≥60 years of age (P=0.002). The median time to explant for SVD was 10.7 years. SVD presented mostly as acute, severe aortic insufficiency attributed to leaflet tear (77.8%). The independent risk factors for reoperation for SVD were age <60 years (P=0.001) and dyslipidemia (P=0.02).
CONCLUSIONS: Aortic valve replacement with the Freestyle bioprosthesis in a subcoronary position provides good long-term clinical and echocardiographic outcomes for patients >60 years of age. Severe aortic insufficiency with leaflet tear is the major mode of SVD leading to reoperation in these patients.
METHODS AND RESULTS: Between 1993 and 2011, 430 patients underwent primary aortic valve replacement with a Freestyle bioprosthesis in the subcoronary position. Mean age was 68.2 ± 8.2 years. All of the clinical and echocardiographic data were collected prospectively. Mean overall follow-up was 9.1 ± 4.4 years and was complete in all of the patients. In-hospital mortality was 3.5% (n=15). Overall, 10- and 15-year survival were 60.7% and 35.0%, respectively. Fifty-one patients required reoperation during follow-up, including 27 for structural valve deterioration (SVD). Overall, freedom from reoperation was 91.0% and 75.0% at 10 and 15 years, whereas freedom from reoperation for SVD was 95.9% and 82.3%, respectively. At 10 and 15 years, freedom from reoperation for SVD was 94.0% and 62.6% for patients <60 years of age and 96.3% and 88.4% for patients ≥60 years of age (P=0.002). The median time to explant for SVD was 10.7 years. SVD presented mostly as acute, severe aortic insufficiency attributed to leaflet tear (77.8%). The independent risk factors for reoperation for SVD were age <60 years (P=0.001) and dyslipidemia (P=0.02).
CONCLUSIONS: Aortic valve replacement with the Freestyle bioprosthesis in a subcoronary position provides good long-term clinical and echocardiographic outcomes for patients >60 years of age. Severe aortic insufficiency with leaflet tear is the major mode of SVD leading to reoperation in these patients.
Предметные рубрики MeSH
Age FactorsAgedAortic ValveAortic Valve InsufficiencyBioprosthesisDisease-Free SurvivalEquipment FailureFemaleFollow-Up StudiesHeart Valve ProsthesisHospital MortalityHumansKaplan-Meier EstimateMaleMiddle AgedPostoperative ComplicationsProspective StudiesProsthesis DesignReoperationTreatment OutcomeUltrasonography