[Accuracy of Setup in Deep-inspiration Breath-hold Irradiation Following Left Mastectomy Using Surface Imaging Guidance].
Аннотация на английскомСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Tanoue Minori, Matsushita Norimasa, Sasaki Makoto, Fujimoto Takahiro, Nakata Manabu, Ono Yuka, Yoshimura Michio, Mizowaki Takashi
Аннотация
PURPOSE: To evaluate the setup accuracy of surface-guided postmastectomy radiotherapy (PMRT) under deep inspiration breath hold (DIBH).
METHODS: Fourteen patients with left-sided breast cancer who underwent PMRT under DIBH were enrolled in this study. The setup error was defined as the difference in the position of clinical target volume between the planning computed tomography (CT) and the cone-beam CT acquired after surface-guided setup.
RESULTS: The mean±standard deviation of setup error was 0.2±2.1 mm, -0.1±3.5 mm, and 0.4±2.1 mm in the anterior-posterior, superior-inferior, and right-left directions, respectively, and -0.1±0.8°, 0.5±0.9°, and 0.9±1.0° in the Yaw, Roll, and Pitch angles.
CONCLUSION: The setup accuracy of PMRT under DIBH using a surface-guided system was within 5 mm in 82% of all treatment fractions.
METHODS: Fourteen patients with left-sided breast cancer who underwent PMRT under DIBH were enrolled in this study. The setup error was defined as the difference in the position of clinical target volume between the planning computed tomography (CT) and the cone-beam CT acquired after surface-guided setup.
RESULTS: The mean±standard deviation of setup error was 0.2±2.1 mm, -0.1±3.5 mm, and 0.4±2.1 mm in the anterior-posterior, superior-inferior, and right-left directions, respectively, and -0.1±0.8°, 0.5±0.9°, and 0.9±1.0° in the Yaw, Roll, and Pitch angles.
CONCLUSION: The setup accuracy of PMRT under DIBH using a surface-guided system was within 5 mm in 82% of all treatment fractions.