Evaluation of the effect of metal stents on dose perturbation in the carbon beam irradiation field.

Miyasaka Yuya, Ishizawa Tetsuya, Nawa Yoshihito, Souda Hikaru, Kawashiro Shohei, Chai Hongbo, Ishizawa Miyu, Sato Hiraku, Iwai Takeo

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

PROPOSE: Carbon ion therapy is indicated for cases in which stents have been inserted, such as bile ducts, but the effect of metal stents on carbon ion therapy is unclear. In this study, the dose perturbation of carbon ion therapy caused by metallic bile duct stents was evaluated by dosimetry.

MATERIALS AND

METHODS

Five different types of metal stents (EGIS Double Bear Biliary Stent, EGIS Single Bear Covered Biliary Stent, BileRush Selective, Niti-S Less Shortening D-type Stent, and ZEO Stent V) were placed between solid phantoms and Gafchromic film were placed around the stents and at each depth. The phantom was irradiated with a Carbon ion Beam to form a 5 cm × 5 cm × 5 cm spread-out Brag peak. The dose change due to the stent was evaluated by comparing the film values of similar irradiations without the stent in place.

RESULT: The results showed that the dose perturbation with and without the stent was <10%. A dose reduction area along the shape of the stent appeared at the end of the irradiation field.

CONCLUSION

The effect of metal stents on the carbon dose distribution was assumed to be comparable to that of x-rays and protons.

Для цитирования: Miyasaka Yuya, Ishizawa Tetsuya, Nawa Yoshihito, Souda Hikaru, Kawashiro Shohei, Chai Hongbo, Ishizawa Miyu, Sato Hiraku, Iwai Takeo. Evaluation of the effect of metal stents on dose perturbation in the carbon beam irradiation field.. Journal of applied clinical medical physics. 2025. DOI: 10.1002/acm2.70042.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.