Evaluation of the dose reduction effect of crystalline lens exposure in cone-beam computed tomography with bismuth eye shield for image-guided radiation therapy: An anthropomorphic phantom study.

Yoshida Tatsuya, Sasaki Koji, Kawasaki Yoshiyuki, Hayakawa Tomoki, Kawadai Toshiyuki, Shibasaki Takako

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

This study aimed to evaluate the dose-exposure reduction effect of a crystalline lens with a bismuth eye shield using cone-beam computed tomography (CBCT) for head image-guided radiation therapy. The ocular surface dose of the head phantom (THRA-1) is defined as a crystalline lens exposure dose and is measured using a radiophotoluminescence dosimeter (RPLD, GD-352 M) with and without an eye shield (CT eye shield) while moving the head phantom from the reference position that is set at the center of the head in either the X or Z direction from -5 to +5 cm. The exposure doses were measured thrice at each movement position. The crystalline lens exposure doses at the reference position were 0.896 ± 0.024 mGy and 0.892 ± 0.016 mGy for the right and left sides, respectively. The exposure doses at the position where the head phantom was moved 5 cm in the -Z direction from the reference position were 2.812 ± 0.053 mGy and 2.576 ± 0.038 mGy for the right and left sides, respectively, with the highest doses at all movement positions. The crystalline lens exposure doses were reduced to 1.909 ± 0.046 mGy and 1.768 ± 0.043 mGy for the right and left sides with an eye shield in this position, causing an exposure dose reduction rate of -32% and -31%, respectively. The crystalline lens exposure dose reduction rate was approximately 10%-15% in the movement directions, except for the -Z direction. Head CBCT with an eye shield effectively reduced the crystalline lens exposure dose when the CBCT isocenter was set close to the eye. Head CBCT using an eye shield is a useful method that reduces the crystalline lens exposure dose.

Для цитирования: Yoshida Tatsuya, Sasaki Koji, Kawasaki Yoshiyuki, Hayakawa Tomoki, Kawadai Toshiyuki, Shibasaki Takako. Evaluation of the dose reduction effect of crystalline lens exposure in cone-beam computed tomography with bismuth eye shield for image-guided radiation therapy: An anthropomorphic phantom study.. Journal of applied clinical medical physics. 2025. DOI: 10.1002/acm2.70024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.