Are the same tapes really the same? Ultrasound study of laser-cut and mechanically cut TVT-O post-operative behavior.

Rusavy Zdenek, Masata Jaromir, Svabik Kamil, Hubka Petr, Zvara Karel, Martan Alois

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION AND HYPOTHESIS: TVT-O production has been modified to laser cutting from mechanical cutting. We compared the behavior of laser and mechanically cut tension-free vaginal tape-obturator (TVT-O) using ultrasound at various time points after surgery.

METHODS

This is a retrospective analysis of clinical and ultrasound data from two previously reported randomized controlled trials with TVT-O. Behavior of mechanically cut TVT-O implanted in January 2007 to November 2009 and laser-cut TVT-O implanted in May 2010 to May 2012 was assessed by ultrasound at day 1, the 2nd week, the 3rd month, and the 1st and 2nd years post-operatively. Bladder neck and tape margins positions were described by coordinates in the orthogonal system calculated from polar coordinates. Tape mobility was measured as a change in the upper and lower tape margin position from rest to maximal Valsalva. Comparison of 2-year subjective and objective surgery outcomes was also performed.

RESULTS

In total, 68 mechanically cut and 50 laser-cut TVT-Os were implanted. Follow-up data were available from 49 and 45 women respectively. No differences in any baseline characteristics or bladder neck mobility were observed. Significantly lower tape mobility was observed on day 1 and week 2 after mechanically cut TVT-O, although subsequent mobility was comparable to laser-cut TVT-O. The subjective and objective surgery outcomes were comparable.

CONCLUSIONS

Although without clinical significance, early postoperative behavior of the mechanically cut and laser-cut TVT-O tapes differs. The less stiff, mechanically-cut TVT-O loosens within 2 weeks of implantation, whereas the stiffer, laser-cut TVT-O keeps its tension.

Для цитирования: Rusavy Zdenek, Masata Jaromir, Svabik Kamil, Hubka Petr, Zvara Karel, Martan Alois. Are the same tapes really the same? Ultrasound study of laser-cut and mechanically cut TVT-O post-operative behavior.. International urogynecology journal. 2017. DOI: 10.1007/s00192-017-3516-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.