Anesthetic vaporizer mount malfunction resulting in oxygenation failure after initiating cardiopulmonary bypass: specific recommendations for the pre-bypass checklist.

Gautam Nischal K, Schmitz Michael L, Zabala Luis M, White Michael W, Mckamie Wesley A, Lutz Alyssa, Johnson Charles E

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Modern technologic advances in medicine have allowed commonly used machines to perform safely with very low risk and a high degree of success. To detect or prevent potential malfunctions, professionals routinely perform pre-use checks for equipment such as anesthesia machines and cardiopulmonary bypass (CPB) machines. These machine checklists are not only critical for a safe operation but also have large impacts on outcomes. For example, when malfunctions are encountered that could have potential negative ramifications or adverse outcomes, multi-approach strategies should be used to identify rectifiable causes and find solutions that are practical. This information can be used to promulgate safe practice guidelines. This case report identifies a machine-based contributing factor to precipitous hypoxia on initiation of bypass in one of our patients. After a detailed approach to identify preventable root causes, we made simple additions to our pre-bypass checklist and recommend these changes to other institutions.

Для цитирования: Gautam Nischal K, Schmitz Michael L, Zabala Luis M, White Michael W, Mckamie Wesley A, Lutz Alyssa, Johnson Charles E. Anesthetic vaporizer mount malfunction resulting in oxygenation failure after initiating cardiopulmonary bypass: specific recommendations for the pre-bypass checklist.. The journal of extra-corporeal technology. 2009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.