MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jtcvs.2024.11.041

Медицинская библиотека · The Journal of thoracic and cardiovascular surgery · 2025

Endo-Bentall repair: Early results and feasibility of a physician-constructed Endo-Bentall device.

Ghoreishi Mehrdad, Shah Aakash, Chahal Diljon, Kang Jeanwan, Gupta Anuj, Taylor Bradly S, Toursavadkohi Shahab

СтатьяVideo-Audio MediaПеревод в очереди

Аннотация

OBJECTIVE

We present our experience with the endovascular Bentall procedure (Endo-Bentall) using a modular valve conduit in high-risk patients with aortic root pathologies.

METHODS

The physician-constructed Endo-Bentall device is composed of a self-expanding transcatheter aortic valve (TAVR), thoracic endovascular aortic stent graft (TEVAR), and 2 wire-reinforced fenestrations for coronary artery stenting. The TAVR valve is sutured into an appropriately sized TEVAR graft. The coronary fenestrations are made in the TEVAR graft based on computed tomography angiography measurements. The Endo-Bentall device is then resheathed within the TEVAR delivery system, advanced via femoral access, and deployed during rapid pacing. If persistent endoleak is present after device deployment, the coronary ostia are accessed and stented. No cardiopulmonary support is required.

RESULTS

Five patients with prohibitive risk (mean age, 79 years; 4 with type A dissection and 1 with an aortic root aneurysm) underwent Endo-Bentall repair between November 2022 and October 2023. Severe aortic insufficiency was present in 2 patients preoperatively. The median back table modification time was 120 minutes (range, 55-150 minutes). The Endo-Bentall device was successfully implanted in all patients (100%). Two patients required concomitant total endoarch repair. In-hospital mortality was 0%. No patient experienced postoperative stroke. Three patients needed coronary stenting, and 1 patient required permanent pacemaker insertion. On follow-up computed tomography scan (at a median of 172 days; range, 63-368 days), no patient had a type 1a endoleak. No aortic insufficiency was present on follow-up transthoracic echocardiography.

CONCLUSIONS

Endo-Bentall repair using a single-stage valve-carrying conduit is feasible and effective, with encouraging early outcomes for highly selected patients with aortic root dissection or aneurysm who are considered not candidates for operation. Long-term follow-up data are needed to assess the persistence of the seal and ongoing durability of this novel technique.

Для цитирования: Ghoreishi Mehrdad, Shah Aakash, Chahal Diljon, Kang Jeanwan, Gupta Anuj, Taylor Bradly S, Toursavadkohi Shahab. Endo-Bentall repair: Early results and feasibility of a physician-constructed Endo-Bentall device.. The Journal of thoracic and cardiovascular surgery. 2025. DOI: 10.1016/j.jtcvs.2024.11.041.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.