Robotic Assisted Transcranial Doppler Monitoring in Acute Neurovascular Care: A Feasibility and Safety Study.

Fattorello Salimbeni Alvise, Kulyk Caterina, Favruzzo Francesco, De Rosa Ludovica, Viaro Federica, Pieroni Alessio, Mozzetta Stefano, Vosko Milan R, Baracchini Claudio

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Transcranial color Doppler (TCD) is currently the only noninvasive bedside tool capable of providing real-time information on cerebral hemodynamics. However, being operator dependent, TCD monitoring is not feasible in many institutions. Robotic assisted TCD (ra-TCD) was recently developed to overcome these constraints. The aim of this study was to evaluate the safety and feasibility of cerebral monitoring with a novel ra-TCD in acute neurovascular care.

METHODS

This is a two-center prospective study conducted between August 2021 and February 2022 at Padua University Hospital (Padua, Italy) and Kepler University Hospital (Linz, Austria). Adult patients with conditions impacting cerebral hemodynamics or patients undergoing invasive procedures affecting cerebral hemodynamics were recruited for prolonged monitoring (> 30 min) of the middle cerebral artery with a novel ra-TCD (NovaGuide Intelligent Ultrasound, NeuraSignal, Los Angeles, CA). Manual TCD was also performed for comparison by an experienced operator. Feasibility and safety rates were recorded.

RESULTS

A total of 92 patients (age: mean 68.5 years, range 36-91; sex: male 57 [62%]) were enrolled in the two centers: 54 in Padua, 38 in Linz. The examination was feasible in the majority of patients (85.9%); the head cradle design and its radiopacity hindered its use during carotid endarterectomy and mechanical thrombectomy. Regarding safety, only one patient (1.1%) reported a minor local edema due to prolonged probe pressure. Velocity values were similar between ra-TCD and manual TCD.

CONCLUSIONS

This novel ra-TCD showed an excellent safety and feasibility and proved to be as reliable as manual TCD in detecting blood flow velocities. These findings support its wider use for cerebral hemodynamics monitoring in acute neurovascular care. However, further technical improvements are needed to expand the range of applicable settings.

Для цитирования: Fattorello Salimbeni Alvise, Kulyk Caterina, Favruzzo Francesco, De Rosa Ludovica, Viaro Federica, Pieroni Alessio, Mozzetta Stefano, Vosko Milan R, Baracchini Claudio. Robotic Assisted Transcranial Doppler Monitoring in Acute Neurovascular Care: A Feasibility and Safety Study.. Neurocritical care. 2024. DOI: 10.1007/s12028-024-02121-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.