Head-to-head evaluation of a commercial real-time PCR assay targeting plasmodial species causing human malaria against a composite reference standard.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Hennies Marc Tim, Rohde Holger, May Jürgen, Tanida Konstantin, Tannich Egbert, Frickmann Hagen
Аннотация
BACKGROUND: The investigation was conducted to evaluate a robust commercial multiplexed species- or species complex-specific real-time PCR assay for human malaria applicable with mobile cyclers.
METHODS: A well-characterized sample set of blood samples from travelers with high pre-test probability of being infected with plasmodial species was assessed with the Plasmodium qPCR panel (Bioeksen, Istanbul, Turkey) against a composite reference standard comprising microscopy following the protocols by World Health Organization in combination with four nucleic acid amplification tests for malaria.
RESULTS: From 1,020 included samples, composite reference standard testing identified 248 (24.3%) as positive for plasmodial DNA, comprising 217x Plasmodium falciparum, 14x Plasmodium vivax, 8x Plasmodium ovale complex, 6x Plasmodium malariae and 3 samples containing DNA of both P. falciparum and P. vivax. Diagnostic accuracy of the evaluated PCR in terms of sensitivity and specificity was 96.8% (240/248) and 99.5% (768/772). The resulting positive and negative predictive values were 98.4% (240/244) and 99.0% (768/776). False-negative results were associated with either submicroscopic parasitemia or parasitemia at the threshold of microscopic detection.
CONCLUSION: The assessed multiplex real-time PCR assay showed similar performance characteristics like comparable diagnostic assays and thus proved to be suitable for the differentiation of plasmodia in the non-endemic setting.
METHODS: A well-characterized sample set of blood samples from travelers with high pre-test probability of being infected with plasmodial species was assessed with the Plasmodium qPCR panel (Bioeksen, Istanbul, Turkey) against a composite reference standard comprising microscopy following the protocols by World Health Organization in combination with four nucleic acid amplification tests for malaria.
RESULTS: From 1,020 included samples, composite reference standard testing identified 248 (24.3%) as positive for plasmodial DNA, comprising 217x Plasmodium falciparum, 14x Plasmodium vivax, 8x Plasmodium ovale complex, 6x Plasmodium malariae and 3 samples containing DNA of both P. falciparum and P. vivax. Diagnostic accuracy of the evaluated PCR in terms of sensitivity and specificity was 96.8% (240/248) and 99.5% (768/772). The resulting positive and negative predictive values were 98.4% (240/244) and 99.0% (768/776). False-negative results were associated with either submicroscopic parasitemia or parasitemia at the threshold of microscopic detection.
CONCLUSION: The assessed multiplex real-time PCR assay showed similar performance characteristics like comparable diagnostic assays and thus proved to be suitable for the differentiation of plasmodia in the non-endemic setting.