m+Medelevator.RU
← Поиск публикаций
Acta otorrinolaringologica espanola · 2018

Hearing outcomes of the active bone conduction system Bonebridge® in conductive or mixed hearing loss.

Оценочное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Carnevale Claudio, Til-Pérez Guillermo, Arancibia-Tagle Diego J, Tomás-Barberán Manuel D, Sarría-Echegaray Pedro L

Аннотация

OBJECTIVE: The active transcutaneous bone conduction implant Bonebridge®, is indicated for patients affected by bilateral conductive/mixed hearing loss or unilateral sensorineural hearing loss, showing hearing outcomes similar to other percutaneous bone conduction implants, but with a lower rate of complications. The aim of this study was to analyze the hearing outcomes in a series of 26 patients affected by conductive or mixed hearing loss and treated with Bonebridge®.

METHODS: 26 of 30 patients implanted with Bonebridge® between October 2012 and May 2017, were included in the study. We compared the air conduction thresholds at the frequencies 500, 1000, 2000, 3000, 4000Hz, the SRT50% and the percentage of correct answers at an intensity of 50dB with and without the implant.

RESULTS: "Pure tone average" with the implant was 34.91dB showing an average gain of 33.46dB. Average SRT 50% with the implant was 34.33dB, whereas before the surgery no patient achieved 50% of correct answers at a sound intensity of 50dB. The percentage of correct answers at 50dB changed from 11% without the implant to 85% with it. We only observed one complication consisting of an extrusion of the implant in a patient with a history of 2 previous rhytidectomies.

CONCLUSIONS: The hearing outcomes obtained in our study are similar to those published in the literature. Bonebridge® represents an excellent alternative in the treatment of conductive or mixed hearing loss, and with a lower rate of complications.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты