Finding a proper "Mate": Comparison of left ventricular assist devices using cerebral oximetry.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Montgomery Morgan L, Stannard Blaine, Dzedzik Siarhei, Lin Hung-Mo, Anyanwu Anelechi, Levin Matthew A, Trinh Muoi
Аннотация
INTRODUCTION: Axial-flow and centrifugal-flow left ventricular assist devices (LVAD) have been utilized in the management of heart failure, but it remains unknown whether these devices differ in end-organ perfusion. Our goal was to evaluate the association between device type and regional cerebral oxygen saturation (rSO2), and determine if this confers any benefit in short-term postoperative outcomes.
METHODS: Adult patients who underwent primary LVAD implantation at our institution from 2014 to 2019 were retrospectively analyzed. Patients were stratified into axial-flow and centrifugal-flow groups. Intraoperative rSO2 readings were used to calculate the change in mean rSO2 from pre- to post-bypass. Multivariable modeling was performed to compare delta rSO2 between groups, and to analyze the association between LVAD type and postoperative outcomes.
RESULTS: There were 152 patients included, of which 76 had an axial-flow device and 76 had a centrifugal-flow device implanted. The rSO2 level increased from pre-bypass to post-bypass on average 3.5% (CI: 2.1 to 5.0) for the axial group compared to 0.1% (CI: -1.2 to 1.4) for the centrifugal group, which was a significant difference (β = -2.22, CI: -4.21 to -0.32, p = 0.022). Axial devices approached significance for lower odds of postoperative complications (OR: 0.35, CI: 0.11 to 1.06, p = 0.063), and were associated with significantly shorter ICU LOS (β = -0.36, CI: -0.60 to -0.11, p = 0.004).
CONCLUSION: Axial devices resulted in a greater increase in rSO2 than centrifugal pumps after separation from CPB. Further investigation is warranted to evaluate the effect of LVAD selection on long-term end-organ perfusion and subsequent patient outcomes.
METHODS: Adult patients who underwent primary LVAD implantation at our institution from 2014 to 2019 were retrospectively analyzed. Patients were stratified into axial-flow and centrifugal-flow groups. Intraoperative rSO2 readings were used to calculate the change in mean rSO2 from pre- to post-bypass. Multivariable modeling was performed to compare delta rSO2 between groups, and to analyze the association between LVAD type and postoperative outcomes.
RESULTS: There were 152 patients included, of which 76 had an axial-flow device and 76 had a centrifugal-flow device implanted. The rSO2 level increased from pre-bypass to post-bypass on average 3.5% (CI: 2.1 to 5.0) for the axial group compared to 0.1% (CI: -1.2 to 1.4) for the centrifugal group, which was a significant difference (β = -2.22, CI: -4.21 to -0.32, p = 0.022). Axial devices approached significance for lower odds of postoperative complications (OR: 0.35, CI: 0.11 to 1.06, p = 0.063), and were associated with significantly shorter ICU LOS (β = -0.36, CI: -0.60 to -0.11, p = 0.004).
CONCLUSION: Axial devices resulted in a greater increase in rSO2 than centrifugal pumps after separation from CPB. Further investigation is warranted to evaluate the effect of LVAD selection on long-term end-organ perfusion and subsequent patient outcomes.