Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA-Compact-Max analyser.
Аннотация
INTRODUCTION
Haemolysis, icterus and lipaemia (HIL) may affect haemostasis test results. This may be influenced by the level of interfering substance and the reagents and end-point detection system used.
METHODS
We assessed the influence of HIL on prothrombin time, activated partial thromboplastin time and fibrinogen assay using a viscosity-based detection analyser.
RESULTS
Spontaneous haemolysis that occurred during sample collection and processing had no effect on PT with either a rabbit tissue factor extract or recombinant human tissue factor reagents. In contrast, addition of mechanically haemolysed cells impacted on PT for the highest haemoglobin concentration. For APTTs determined with STA®-Cephascreen® reagent, there was no significant difference between results in haemolysed and nonhaemolysed samples. For the other two reagents studied, APTTs were statistically significantly shorter in haemolysed samples compared with nonhaemolysed samples. This bias was clinically significant only for STA®-PTT Automate. For all three APTT reagents, the impact of haemolysis was sufficient to impact patient management decisions, and in some samples, the effects of lipaemia and icterus were not clinically significant.
CONCLUSION
Overall, our results confirm that PT and fibrinogen were not clinically significantly affected by HIL. The APTTs of some haemolysed samples were falsely normal. Haemolysed samples for APTT determination should be rejected.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.