DOI 10.1111/ijlh.12498

Медицинская библиотека · International journal of laboratory hematology · 2016

Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA-Compact-Max analyser.

Woolley A, Golmard J-L, Kitchen S

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Haemolysis, icterus and lipaemia (HIL) may affect haemostasis test results. This may be influenced by the level of interfering substance and the reagents and end-point detection system used.

METHODS

We assessed the influence of HIL on prothrombin time, activated partial thromboplastin time and fibrinogen assay using a viscosity-based detection analyser.

RESULTS

Spontaneous haemolysis that occurred during sample collection and processing had no effect on PT with either a rabbit tissue factor extract or recombinant human tissue factor reagents. In contrast, addition of mechanically haemolysed cells impacted on PT for the highest haemoglobin concentration. For APTTs determined with STA®-Cephascreen® reagent, there was no significant difference between results in haemolysed and nonhaemolysed samples. For the other two reagents studied, APTTs were statistically significantly shorter in haemolysed samples compared with nonhaemolysed samples. This bias was clinically significant only for STA®-PTT Automate. For all three APTT reagents, the impact of haemolysis was sufficient to impact patient management decisions, and in some samples, the effects of lipaemia and icterus were not clinically significant.

CONCLUSION

Overall, our results confirm that PT and fibrinogen were not clinically significantly affected by HIL. The APTTs of some haemolysed samples were falsely normal. Haemolysed samples for APTT determination should be rejected.

Для цитирования: Woolley A, Golmard J-L, Kitchen S. Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA-Compact-Max analyser.. International journal of laboratory hematology. 2016. DOI: 10.1111/ijlh.12498.