DOI 10.1055/a-1010-6052

Медицинская библиотека · Ultraschall in der Medizin (Stuttgart, Germany : 1980) · 2019

Establishing Reliability Criteria for Liver ElastPQ Shear Wave Elastography (ElastPQ-SWE): Comparison Between 10, 5 and 3 Measurements.

Roccarina Davide, Iogna Prat Laura, Buzzetti Elena, Guerrero Misas Marta, Aricó Francesco Marcello, Saffioti Francesca, Rosselli Matteo, Pinzani Massimo, Marshall Aileen, Thorburn Douglas, Tsochatzis Emmanuel

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: ElastPQ is a new elastography technique for non-invasive liver fibrosis staging. However, it does not have validated reliability criteria. We tested the reliability of a different number of measurements in patients with chronic liver disease and explored whether the application of quality criteria improves the diagnostic performance.

MATERIALS AND

METHODS

All patients underwent liver stiffness assessment (LSM) with ElastPQ and Fibroscan (F-TE). The mean, median, standard deviation (SD) and interquartile range (IQR) of 10, 5 and 3 measurements were retrospectively collected for each patient and compared to each other. Liver histology was available in a subset of patients.

RESULTS

Overall, 400 patients met the inclusion criteria. Non-alcoholic fatty liver disease (NAFLD) was the most represented etiology (75 %), followed by primary sclerosing cholangitis (PSC), primary biliary cholangitis (PBC) and autoimmune hepatitis (AIH). The correlation of medians was significantly better between 10 and 5 measurements than between 10 and 3. The difference of medians was significant only in the comparison between 10 and 3 measurements. The correlation between ElastPQ and F-TE was equally good for 10 and 5 measurements and significantly improved after an IQR/median ≤ 30 % was applied. The diagnostic performance of ElastPQ was better with the median value of 10 and 5 measurements and improved if LSM values were obtained with IQR/M ≤ 30 %.

CONCLUSION

The median value of 5 valid LSMs suffices for the reliable estimation of liver stiffness using ElastPQ. The quality criterion of IQR/M ≤ 30 % should also be followed when using this technique.

Для цитирования: Roccarina Davide, Iogna Prat Laura, Buzzetti Elena, Guerrero Misas Marta, Aricó Francesco Marcello, Saffioti Francesca, Rosselli Matteo, Pinzani Massimo, Marshall Aileen, Thorburn Douglas, Tsochatzis Emmanuel. Establishing Reliability Criteria for Liver ElastPQ Shear Wave Elastography (ElastPQ-SWE): Comparison Between 10, 5 and 3 Measurements.. Ultraschall in der Medizin (Stuttgart, Germany : 1980). 2019. DOI: 10.1055/a-1010-6052.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.