MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/09638280701615246

Медицинская библиотека · Disability and rehabilitation · 2008

Patient-driven loop control for hand function restoration in a non-invasive functional electrical stimulation system.

Chiou Ying-Han, Luh Jer-Junn, Chen Shih-Ching, Chen Yu-Luen, Lai Jin-Shin, Kuo Te-Son

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: In this study, a patient-driven loop control in a non-invasive functional electrical stimulation (FES) system was designed to restore hand function of stroke patients with their residual capabilities.

METHOD: With this patient-driven loop control, patients use the electromyographic (EMG) signals from their voluntary controlled muscles in affected limbs to adjust stimulus parameters of the system. A special designed FES system generated electrical stimuli to excite the paralyzed muscles through surface electrodes on the basis of the control command from the residual myoelectric signals. EMG signals were also served as the trigger and the adjustment of stimulus parameters and thereby adding versatility of the FES system. Four stroke patients were recruited in the experiment to validate our system.

RESULTS

The experimental results showed that hemiplegics could successfully control the system to restore their lost hand functions with the strategy of patient-driven loop control (the average estimated success rate was 77.5% with the tasks of cylindrical grasp and lateral pinch); and further, they would benefit by using the residual capabilities to regain their hand functions from the viewpoints of rehabilitation and psychology.

CONCLUSION

According to the experiment results, this patient-driven loop control can be beneficial for hemiplegics to restore their hand functions such as cylindrical grasp and lateral pinch. The control strategy of this study has the potential to be employed not only in the FES system but also in other assistive devices.

Для цитирования: Chiou Ying-Han, Luh Jer-Junn, Chen Shih-Ching, Chen Yu-Luen, Lai Jin-Shin, Kuo Te-Son. Patient-driven loop control for hand function restoration in a non-invasive functional electrical stimulation system.. Disability and rehabilitation. 2008. DOI: 10.1080/09638280701615246.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.