In vivo calibration of the T2* cardiovascular magnetic resonance method at 1.5 T for estimation of cardiac iron in a minipig model of transfusional iron overload.

Jensen Peter Diedrich, Nielsen Asbjørn Haaning, Simonsen Carsten Wiberg, Baandrup Ulrik Thorngren, Jensen Svend Eggert, Bøgsted Martin, Magnusdottir Sigridur Olga, Jensen Anne Birthe Helweg, Kjaergaard Benedict

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Non-invasive estimation of the cardiac iron concentration (CIC) by T2* cardiovascular magnetic resonance (CMR) has been validated repeatedly and is in widespread clinical use. However, calibration data are limited, and mostly from post-mortem studies. In the present study, we performed an in vivo calibration in a dextran-iron loaded minipig model.

METHODS

R2* (= 1/T2*) was assessed in vivo by 1.5 T CMR in the cardiac septum. Chemical CIC was assessed by inductively coupled plasma-optical emission spectroscopy in endomyocardial catheter biopsies (EMBs) from cardiac septum taken during follow up of 11 minipigs on dextran-iron loading, and also in full-wall biopsies from cardiac septum, taken post-mortem in another 16  minipigs, after completed iron loading.

RESULTS

A strong correlation could be demonstrated between chemical CIC in 55 EMBs and parallel cardiac T2* (Spearman rank correlation coefficient 0.72, P < 0.001). Regression analysis led to [CIC] = (R2* - 17.16)/41.12 for the calibration equation with CIC in mg/g dry weight and R2* in Hz. An even stronger correlation was found, when chemical CIC was measured by full-wall biopsies from cardiac septum, taken immediately after euthanasia, in connection with the last CMR session after finished iron loading (Spearman rank correlation coefficient 0.95 (P < 0.001). Regression analysis led to the calibration equation [CIC] = (R2* - 17.2)/31.8.

CONCLUSIONS

Calibration of cardiac T2* by EMBs is possible in the minipig model but is less accurate than by full-wall biopsies. Likely explanations are sampling error, variable content of non-iron containing tissue and smaller biopsies, when using catheter biopsies. The results further validate the CMR T2* technique for estimation of cardiac iron in conditions with iron overload and add to the limited calibration data published earlier.

Для цитирования: Jensen Peter Diedrich, Nielsen Asbjørn Haaning, Simonsen Carsten Wiberg, Baandrup Ulrik Thorngren, Jensen Svend Eggert, Bøgsted Martin, Magnusdottir Sigridur Olga, Jensen Anne Birthe Helweg, Kjaergaard Benedict. In vivo calibration of the T2* cardiovascular magnetic resonance method at 1.5 T for estimation of cardiac iron in a minipig model of transfusional iron overload.. Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance. 2021. DOI: 10.1186/s12968-021-00715-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.