MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.28476

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2019

The initial U.S. experience with the Tempo active fixation temporary pacing lead in structural heart interventions.

Nazif Tamim M, Chen Shmuel, Codner Pablo, Grossman Paul M, Menees Daniel S, Sanchez Carlos E, Yakubov Steven J, White Jonathan, Kapadia Samir, Whisenant Brian K, Forrest John K, Krishnaswamy Amar, Arshi Arash, Orford James L, Leon Martin B, Dizon José M, Kodali Susheel K, Chetcuti Stanley J

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This multicenter retrospective study of the initial U.S. experience evaluated the safety and efficacy of temporary cardiac pacing with the Tempo® Temporary Pacing Lead.

BACKGROUND

Despite increasing use of temporary cardiac pacing with the rapid growth of structural heart procedures, temporary pacing leads have not significantly improved. The Tempo lead is a new temporary pacing lead with a soft tip intended to minimize the risk of perforation and a novel active fixation mechanism designed to enhance lead stability.

METHODS

Data from 269 consecutive structural heart procedures were collected. Outcomes included device safety (absence of clinically significant cardiac perforation, new pericardial effusion, or sustained ventricular arrhythmia) and efficacy (clinically acceptable pacing thresholds with successful pace capture throughout the index procedure). Postprocedure practices and sustained lead performance were also analyzed.

RESULTS

The Tempo lead was successfully positioned in the right ventricle and achieved pacing in 264 of 269 patients (98.1%). Two patients (0.8%) experienced loss of pace capture. Procedural mean pace capture threshold (PCT) was 0.7 ± 0.8 mA. There were no clinically significant perforations, pericardial effusions, or sustained device-related arrhythmias. The Tempo lead was left in place postprocedure in 189 patients (71.6%) for mean duration of 43.3 ± 0.7 hr (range 2.5-221.3 hr) with final PCT of 0.84 ± 1.04 mA (n = 80). Of these patients, 84.1% mobilized out of bed with no lead dislodgment.

CONCLUSION

The Tempo lead is safe and effective for temporary cardiac pacing for structural heart procedures, provides stable peri and postprocedural pacing and allows mobilization of patients who require temporary pacing leads.

Для цитирования: Nazif Tamim M, Chen Shmuel, Codner Pablo, Grossman Paul M, Menees Daniel S, Sanchez Carlos E, Yakubov Steven J, White Jonathan, Kapadia Samir, Whisenant Brian K, Forrest John K, Krishnaswamy Amar, Arshi Arash, Orford James L, Leon Martin B, Dizon José M, Kodali Susheel K, Chetcuti Stanley J. The initial U.S. experience with the Tempo active fixation temporary pacing lead in structural heart interventions.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2019. DOI: 10.1002/ccd.28476.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.