Five commercial 'depth of anaesthesia' monitors provide discordant clinical recommendations in response to identical emergence-like EEG signals.

Hight Darren, Kreuzer Matthias, Ugen Gesar, Schuller Peter, Stüber Frank, Sleigh Jamie, Kaiser Heiko A

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

'Depth of anaesthesia' monitors claim to measure hypnotic depth during general anaesthesia from the EEG, and clinicians could reasonably expect agreement between monitors if presented with the same EEG signal. We took 52 EEG signals showing intraoperative patterns of diminished anaesthesia, similar to those that occur during emergence (after surgery) and subjected them to analysis by five commercially available monitors.

METHODS

We compared five monitors (BIS, Entropy-SE, Narcotrend, qCON, and Sedline) to see if index values remained within, or moved out of, each monitors' recommended index range for general anaesthesia for at least 2 min during a period of supposed lighter anaesthesia, as observed by changes in the EEG spectrogram obtained in a previous study.

RESULTS

Of the 52 cases, 27 (52%) had at least one monitor warning of potentially inadequate hypnosis (index above range) and 16 of the 52 cases (31%) had at least one monitor signifying excessive hypnotic depth (index below clinical range). Of the 52 cases, only 16 (31%) showed concordance between all five monitors. Nineteen cases (36%) had one monitor discordant compared with the remaining four, and 17 cases (33%) had two monitors in disagreement with the remaining three.

CONCLUSIONS

Many clinical providers still rely on index values and manufacturer's recommended ranges for titration decision making. That two-thirds of cases showed discordant recommendations given identical EEG data, and that one-third signified excessive hypnotic depth where the EEG would suggest a lighter hypnotic state, emphasizes the importance of personalised EEG interpretation as an essential clinical skill.

Для цитирования: Hight Darren, Kreuzer Matthias, Ugen Gesar, Schuller Peter, Stüber Frank, Sleigh Jamie, Kaiser Heiko A. Five commercial 'depth of anaesthesia' monitors provide discordant clinical recommendations in response to identical emergence-like EEG signals.. British journal of anaesthesia. 2023. DOI: 10.1016/j.bja.2022.12.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.