New backstrap vagus electrode for continuous intraoperative neuromonitoring in thyroid surgery.
Аннотация
INTRODUCTION
Continuous intraoperative neuromonitoring (CIONM) via vagal nerve stimulation allows real-time surveillance of the recurrent laryngeal nerve during thyroid surgery. However, for effective CIONM, subtle changes in recurrent laryngeal nerve conductivity have to be detected. A newly developed stimulation electrode that provides stable nerve stimulation and safe application is presented.
METHODS
For electrode validation, current distribution was simulated with the finite element method. Mechanical characteristics were assessed through bench testing. Clinical evaluation was initiated with 11 thyroid surgeries.
RESULTS
Experimental and clinical results led to the development of a tripolar gold/polyimide electrode mounted onto a backstrap-shaped silicone body. It facilitated rapid electrode implantation and extraction (median implantation time 4 ± 19 seconds). Peak extraction force was 570 mN. Median supramaximal stimulation currents were 2.00 ± 0.95 mA and resulted in reliable electromyogram responses (median 3.1 ± 3.0 mV). No intraoperative electrode dislocations occurred, and no postoperative nerve palsy was observed.
CONCLUSION
The new backstrap vagal stimulation electrode meets the requirements for reliable CIONM.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.