Clinica chimica acta; international journal of clinical chemistry · 2015
A clinical and laboratory approach used to elucidate discordant results of high-sensitivity troponin T and troponin I.
Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Franeková Janka, Bláha Martin, Bělohoubek Jiří, Kotrbatá Markéta, Sečník Peter, Kubíček Zdenek, Kettner Jiří, Jabor Antonín
Аннотация
BACKGROUND: Careful interpretation of discordant results in high-sensitivity troponin measurements is necessary in cases of suspect immunoassay interferences. We describe several procedures taken in a case of a polymorbid patient with chest pain, without clear evidence of myocardial necrosis and with increased high-sensitivity cardiac troponin T (hs-cTnT). We checked the Vafaie's algorithm for the evaluation of suspect interference in troponin measurements.
METHODS: We conducted a case report analysis, additional measurements, a dilution test and pretreatment of plasma with blocking agents.
RESULTS: Concentration of hs-cTnT (99 th percentile of "Работает" population 14 ng/L) increased from 120.1 ng/L to 280.4 ng/L during an 8-month period and decreased to 216.3 ng/L during the following month with repeatedly negative troponin I (TnI), hs-cTnI, myoglobin and creatine kinase MB (CK-MB). Suspected false positivity of hs-cTnT was further confirmed by treatment of plasma with an antiheterophile blocking agent (hs-cTnT before treatment 280.4 ng/L, after 16.53/16.23 ng/L). This outcome was further confirmed by the manufacturer's experiments.
CONCLUSIONS: The false-positive results of hs-cTnT were caused by the presence of extremely rare high molecular weight protein, presumably IgM, most likely HAMA (human anti-mouse antibody). Only the pre-treatment of plasma with a blocking agent provided a reliable indication of the interference. Cooperation among clinicians, laboratory personnel and the manufacturer is essential.
METHODS: We conducted a case report analysis, additional measurements, a dilution test and pretreatment of plasma with blocking agents.
RESULTS: Concentration of hs-cTnT (99 th percentile of "Работает" population 14 ng/L) increased from 120.1 ng/L to 280.4 ng/L during an 8-month period and decreased to 216.3 ng/L during the following month with repeatedly negative troponin I (TnI), hs-cTnI, myoglobin and creatine kinase MB (CK-MB). Suspected false positivity of hs-cTnT was further confirmed by treatment of plasma with an antiheterophile blocking agent (hs-cTnT before treatment 280.4 ng/L, after 16.53/16.23 ng/L). This outcome was further confirmed by the manufacturer's experiments.
CONCLUSIONS: The false-positive results of hs-cTnT were caused by the presence of extremely rare high molecular weight protein, presumably IgM, most likely HAMA (human anti-mouse antibody). Only the pre-treatment of plasma with a blocking agent provided a reliable indication of the interference. Cooperation among clinicians, laboratory personnel and the manufacturer is essential.