DOI 10.1016/j.cca.2015.03.046

Медицинская библиотека · Clinica chimica acta; international journal of clinical chemistry · 2015

A clinical and laboratory approach used to elucidate discordant results of high-sensitivity troponin T and troponin I.

Franeková Janka, Bláha Martin, Bělohoubek Jiří, Kotrbatá Markéta, Sečník Peter, Kubíček Zdenek, Kettner Jiří, Jabor Antonín

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Careful interpretation of discordant results in high-sensitivity troponin measurements is necessary in cases of suspect immunoassay interferences. We describe several procedures taken in a case of a polymorbid patient with chest pain, without clear evidence of myocardial necrosis and with increased high-sensitivity cardiac troponin T (hs-cTnT). We checked the Vafaie's algorithm for the evaluation of suspect interference in troponin measurements.

METHODS

We conducted a case report analysis, additional measurements, a dilution test and pretreatment of plasma with blocking agents.

RESULTS

Concentration of hs-cTnT (99 th percentile of "Работает" population 14 ng/L) increased from 120.1 ng/L to 280.4 ng/L during an 8-month period and decreased to 216.3 ng/L during the following month with repeatedly negative troponin I (TnI), hs-cTnI, myoglobin and creatine kinase MB (CK-MB). Suspected false positivity of hs-cTnT was further confirmed by treatment of plasma with an antiheterophile blocking agent (hs-cTnT before treatment 280.4 ng/L, after 16.53/16.23 ng/L). This outcome was further confirmed by the manufacturer's experiments.

CONCLUSIONS

The false-positive results of hs-cTnT were caused by the presence of extremely rare high molecular weight protein, presumably IgM, most likely HAMA (human anti-mouse antibody). Only the pre-treatment of plasma with a blocking agent provided a reliable indication of the interference. Cooperation among clinicians, laboratory personnel and the manufacturer is essential.

Для цитирования: Franeková Janka, Bláha Martin, Bělohoubek Jiří, Kotrbatá Markéta, Sečník Peter, Kubíček Zdenek, Kettner Jiří, Jabor Antonín. A clinical and laboratory approach used to elucidate discordant results of high-sensitivity troponin T and troponin I.. Clinica chimica acta; international journal of clinical chemistry. 2015. DOI: 10.1016/j.cca.2015.03.046.