DOI 10.1016/j.acmx.2014.12.001

Медицинская библиотека · Archivos de cardiologia de Mexico · 2015

[New conduction disturbances and pacemaker indications after CoreValve® transcatheter aortic valve replacement. Incidence and follow up in a single center experience].

Aversa Eliana, Muratore Claudio A, Nemesio M Laura, Tentori Maria Cristina, Payaslian Miguel

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Transcatheter aortic valve implantation (TAVI) is currently reserved for patients with symptomatic aortic stenosis and high surgical risk. One major limiting factor related to TAVI procedural complications is conduction abnormalities and the need for permanent pacemaker implantation.

OBJECTIVES

Evaluate the incidence of new conduction disturbances and pacemaker indications in patients with TAVI CoreValve® prosthesis (Medtronic Inc. Minneapolis, Minnesota, United States).

METHODS

We included 28 patients, mean age 80 years. ECG parameters were evaluated previous and after implantation. All patients were monitorized during TAVI. Follow up Holter monitoring was performed at one, 6 and 12 months after the procedure and we also evaluated telemetry of implanted pacemaker.

RESULTS

In previous ECG we found 7 patients had right bundle branch block and 7 patients had left bundle brunch block (LBBB). The post implant ECG showed 7 new LBBB: 3 during valvuloplasty and 4 on the end of it. Six patients required pacemaker implantation for permanent or paroxysmal complete AV block (CAVB). At one year follow up, 3 patients with LBBB during valvuloplasty had a normal ECG, one still had LBBB and one an asymptomatic CAVB found in Holter monitoring.

CONCLUSIONS

Conduction abnormalities are frequent after CoreValve® aortic valve prosthesis implantation. The incidence of new LBBB was 25%. CAVB during or post TAVI require PM implantation. New LBBB may need a closer follow up because in a 3% of the cases it may progress to CAVB.

Для цитирования: Aversa Eliana, Muratore Claudio A, Nemesio M Laura, Tentori Maria Cristina, Payaslian Miguel. [New conduction disturbances and pacemaker indications after CoreValve® transcatheter aortic valve replacement. Incidence and follow up in a single center experience].. Archivos de cardiologia de Mexico. 2015. DOI: 10.1016/j.acmx.2014.12.001.