DOI 10.1007/s00216-022-03984-x

Медицинская библиотека · Analytical and bioanalytical chemistry · 2022

An automated magnetic bead extraction method for measuring plasma metanephrines and 3-methoxytyramine using liquid chromatography tandem mass spectrometry.

Yu Songlin, Zhou Weiyan, Yu Jialei, Li Ming, Zhang Shenyan, Jiang Xu, Wang Huanhong, Ma Xiaoli, Zou Yutong, Wang Danchen, Cheng Qian, Xie Shaowei, Luo Wei, Guo Xiuzhi, Qiu Ling

СтатьяПеревод пока не подготовлен

Аннотация

Liquid chromatography tandem mass spectrometry (LC-MS/MS) is used routinely in clinical diagnostics; however, automating the sample pretreatment is challenging. We established and evaluated an automated method based on the magnetic bead extraction principle (MBE) to measure normetanephrine (NMN), metanephrine (MN), and 3-methoxytyramine (3-MT). The target analytes were extracted, purified, and concentrated using different solvents and chemical bond-modified magnetic beads transferred via a magnetic bar. The linearity, recovery, matrix effect, and precision of MBE were evaluated thoroughly, and compared with traditional solid-phase extraction (SPE) using 131 plasma samples. The chromatography peaks of metanephrines and 3-MT, extracted via MBE, are symmetrical, without interfering peaks. The linearity was excellent with correlation coefficient (r) > 0.99. The MBE exhibited good reproducibility with within-run coefficient variations (CVs) of 1.96-2.00%, 4.06-5.75%, and 3.89-4.90% for MN, NMN, and 3-MT, respectively. The total CVs for MN, NMN, and 3-MT were 1.96-2.80%, 5.12-5.75%, and 5.44-6.27%, respectively. The relative recoveries for MN, NMN, and 3-MT varied between 93.5 and 107.4%, whereas their biases were all within 10%. The results for MN, NMN, and 3-MT extracted via MBE compared with SPE exhibited excellent correlation, with r > 0.99; the mean bias% for MN, NMN, and 3-MT were small (-2.9%, -3.2%, and -3.2%, respectively). In conclusion, the automated MBE method for measuring plasma metanephrines and 3-MT can be applied in future routine clinical diagnostics, and the MBE principle may indicate a new era for LC-MS/MS in clinical application.

Для цитирования: Yu Songlin, Zhou Weiyan, Yu Jialei, Li Ming, Zhang Shenyan, Jiang Xu, Wang Huanhong, Ma Xiaoli, Zou Yutong, Wang Danchen, Cheng Qian, Xie Shaowei, Luo Wei, Guo Xiuzhi, Qiu Ling. An automated magnetic bead extraction method for measuring plasma metanephrines and 3-methoxytyramine using liquid chromatography tandem mass spectrometry.. Analytical and bioanalytical chemistry. 2022. DOI: 10.1007/s00216-022-03984-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.