Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography: effect of multiple B-scan averaging on RNFL measurement.
Аннотация
PURPOSE: To determine the effect of multiple B-scan averaging on retinal nerve fiber layer (RNFL) thickness measurement taken with a spectral-domain optical coherence tomographer (OCT).
METHODS
Twenty-five normal individuals and 29 patients with glaucoma were examined. One eye was selected randomly of each participant for circumpapillary RNFL scanning by the Spectralis OCT (Heidelberg Engineering, Heidelberg, GmbH, Dossenheim, Germany). Three sets of OCT image series comprising 2, 8, and 16 consecutive B-scans were obtained in the same visit. After multiple B-scan averaging, the respective global and sectoral (superonasal, nasal, inferonasal, inferotemporal, temporal, and superotemporal) RNFL thicknesses were compared with linear mixed modeling and analyzed for RNFL measurement variability.
RESULTS
There was no RNFL segmentation failure in all the scans. No significant differences were found in global or sectoral RNFL thicknesses among the image series averaged with different number of B-scan except for the inferonasal sector in the glaucoma group (P=0.036). The intraclass correlation coefficients were all >0.971 and the coefficients of variation were all below 4.33% for both sectoral and global RNFL thicknesses. No significant differences in the image quality score were detected among the image series (P≥0.141).
CONCLUSIONS
Although multiple B-scan averaging is commonly applied to enhance visualization of the retinal layers in OCT images, it may not have a significant impact on global and sectoral RNFL thicknesses measured by the Spectralis OCT.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.