MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s12928-016-0446-5

Медицинская библиотека · Cardiovascular intervention and therapeutics · 2016

External and internal compressions to retrieve a kinked catheter in the right brachial artery.

Nakabayashi Keisuke, Kawakatsu Naomi, Nakazawa Naomi, Suzuki Toshiaki, Okada Hisayuki, Oka Toshiaki

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Kinking of the catheter due to excessive rotation is not a rare complication. However, percutaneous retrieval of a kinked catheter can be difficult. The key to bailout is fixation of the catheter tip. Herein, we present a 78-year-old woman who had this complication during transradial angiography. Retrieval using several previously reported techniques was unsuccessful. We finally retrieved the kinked catheter by fixing the tip of the catheter, using external and internal compressions. The former comprises manual compression on the axillary artery, while the latter comprises deployment of a balloon catheter via another puncture site.

Для цитирования: Nakabayashi Keisuke, Kawakatsu Naomi, Nakazawa Naomi, Suzuki Toshiaki, Okada Hisayuki, Oka Toshiaki. External and internal compressions to retrieve a kinked catheter in the right brachial artery.. Cardiovascular intervention and therapeutics. 2016. DOI: 10.1007/s12928-016-0446-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.