Nanomechanical sensor applied to blood culture pellets: a fast approach to determine the antibiotic susceptibility against agents of bloodstream infections.
Аннотация
OBJECTIVES
The management of bloodstream infection, a life-threatening disease, largely relies on early detection of infecting microorganisms and accurate determination of their antibiotic susceptibility to reduce both mortality and morbidity. Recently we developed a new technique based on atomic force microscopy capable of detecting movements of biologic samples at the nanoscale. Such sensor is able to monitor the response of bacteria to antibiotic's pressure, allowing a fast and versatile susceptibility test. Furthermore, rapid preparation of a bacterial pellet from a positive blood culture can improve downstream characterization of the recovered pathogen as a result of the increased bacterial concentration obtained.
METHODS
Using artificially inoculated blood cultures, we combined these two innovative procedures and validated them in double-blind experiments to determine the susceptibility and resistance of Escherichia coli strains (ATCC 25933 as susceptible and a characterized clinical isolate as resistant strain) towards a selection of antibiotics commonly used in clinical settings.
RESULTS
On the basis of the variance of the sensor movements, we were able to positively discriminate the resistant from the susceptible E. coli strains in 16 of 17 blindly investigated cases. Furthermore, we defined a variance change threshold of 60% that discriminates susceptible from resistant strains.
CONCLUSIONS
By combining the nanomotion sensor with the rapid preparation method of blood culture pellets, we obtained an innovative, rapid and relatively accurate method for antibiotic susceptibility test directly from positive blood culture bottles, without the need for bacterial subculture.