A Dual-Chamber Leadless Pacemaker.

Knops Reinoud E, Reddy Vivek Y, Ip James E, Doshi Rahul, Exner Derek V, Defaye Pascal, Canby Robert, Bongiorni Maria Grazia, Shoda Morio, Hindricks Gerhard, Neužil Petr, Rashtian Mayer, Breeman Karel T N, Nevo Jordan R, Ganz Leonard, Hubbard Chris, Cantillon Daniel J, Aveir DR i2i Study Investigators

СтатьяМногоцентровое исследованиеКлиническое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Single-chamber ventricular leadless pacemakers do not support atrial pacing or consistent atrioventricular synchrony. A dual-chamber leadless pacemaker system consisting of two devices implanted percutaneously, one in the right atrium and one in the right ventricle, would make leadless pacemaker therapy a treatment option for a wider range of indications.

METHODS

We conducted a prospective, multicenter, single-group study to evaluate the safety and performance of a dual-chamber leadless pacemaker system. Patients with a conventional indication for dual-chamber pacing were eligible for participation. The primary safety end point was freedom from complications (i.e., device- or procedure-related serious adverse events) at 90 days. The first primary performance end point was a combination of adequate atrial capture threshold and sensing amplitude at 3 months. The second primary performance end point was at least 70% atrioventricular synchrony at 3 months while the patient was sitting.

RESULTS

Among the 300 patients enrolled, 190 (63.3%) had sinus-node dysfunction and 100 (33.3%) had atrioventricular block as the primary pacing indication. The implantation procedure was successful (i.e., two functioning leadless pacemakers were implanted and had established implant-to-implant communication) in 295 patients (98.3%). A total of 35 device- or procedure-related serious adverse events occurred in 29 patients. The primary safety end point was met in 271 patients (90.3%; 95% confidence interval [CI], 87.0 to 93.7), which exceeded the performance goal of 78% (P<0.001). The first primary performance end point was met in 90.2% of the patients (95% CI, 86.8 to 93.6), which exceeded the performance goal of 82.5% (P<0.001). The mean (±SD) atrial capture threshold was 0.82±0.70 V, and the mean P-wave amplitude was 3.58±1.88 mV. Of the 21 patients (7%) with a P-wave amplitude of less than 1.0 mV, none required device revision for inadequate sensing. At least 70% atrioventricular synchrony was achieved in 97.3% of the patients (95% CI, 95.4 to 99.3), which exceeded the performance goal of 83% (P<0.001).

CONCLUSIONS

The dual-chamber leadless pacemaker system met the primary safety end point and provided atrial pacing and reliable atrioventricular synchrony for 3 months after implantation. (Funded by Abbott Medical; Aveir DR i2i ClinicalTrials.gov number, NCT05252702.).

Для цитирования: Knops Reinoud E, Reddy Vivek Y, Ip James E, Doshi Rahul, Exner Derek V, Defaye Pascal, Canby Robert, Bongiorni Maria Grazia, Shoda Morio, Hindricks Gerhard, Neužil Petr, Rashtian Mayer, Breeman Karel T N, Nevo Jordan R, Ganz Leonard, Hubbard Chris, Cantillon Daniel J, Aveir DR i2i Study Investigators. A Dual-Chamber Leadless Pacemaker.. The New England journal of medicine. 2023. DOI: 10.1056/nejmoa2300080.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.