Automatic Extraction of Myocardial Mass and Volume Using Parametric Images from Dynamic Nongated PET.

Harms Hendrik Johannes, Stubkjær Hansson Nils Henrik, Tolbod Lars Poulsen, Kim Won Yong, Jakobsen Steen, Bouchelouche Kirsten, Wiggers Henrik, Frøkiaer Jørgen, Sörensen Jens

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

UNLABELLED: Dynamic cardiac PET is used to quantify molecular processes in vivo. However, measurements of left ventricular (LV) mass and volume require electrocardiogram-gated PET data. The aim of this study was to explore the feasibility of measuring LV geometry using nongated dynamic cardiac PET.

METHODS

Thirty-five patients with aortic-valve stenosis and 10 healthy controls underwent a 27-min (11)C-acetate PET/CT scan and cardiac MRI (CMR). The controls were scanned twice to assess repeatability. Parametric images of uptake rate K1 and the blood pool were generated from nongated dynamic data. Using software-based structure recognition, the LV wall was automatically segmented from K1 images to derive functional assessments of LV mass (mLV) and wall thickness. End-systolic and end-diastolic volumes were calculated using blood pool images and applied to obtain stroke volume and LV ejection fraction (LVEF). PET measurements were compared with CMR.

RESULTS

High, linear correlations were found for LV mass (r = 0.95), end-systolic volume (r = 0.93), and end-diastolic volume (r = 0.90), and slightly lower correlations were found for stroke volume (r = 0.74), LVEF (r = 0.81), and thickness (r = 0.78). Bland-Altman analyses showed significant differences for mLV and thickness only and an overestimation for LVEF at lower values. Intra- and interobserver correlations were greater than 0.95 for all PET measurements. PET repeatability accuracy in the controls was comparable to CMR.

CONCLUSION

LV mass and volume are accurately and automatically generated from dynamic (11)C-acetate PET without electrocardiogram gating. This method can be incorporated in a standard routine without any additional workload and can, in theory, be extended to other PET tracers.

Для цитирования: Harms Hendrik Johannes, Stubkjær Hansson Nils Henrik, Tolbod Lars Poulsen, Kim Won Yong, Jakobsen Steen, Bouchelouche Kirsten, Wiggers Henrik, Frøkiaer Jørgen, Sörensen Jens. Automatic Extraction of Myocardial Mass and Volume Using Parametric Images from Dynamic Nongated PET.. Journal of nuclear medicine : official publication, Society of Nuclear Medicine. 2016. DOI: 10.2967/jnumed.115.170613.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.