In-vivo dosimetric analysis in total skin electron beam therapy.

Elsayad Khaled, Moustakis Christos, Simonsen Manuela, Bäcker Dagmar, Haverkamp Uwe, Eich Hans Theodor

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: Thermoluminescent dosimetry (TLD) is an important element of total skin electron beam therapy (TSEBT). In this study, we compare radiation dose distributions to provide data for dose variation across anatomic sites.

MATERIALS AND

METHODS

Retrospectively collected data on 85 patients with cutaneous lymphoma or leukemia underwent TSEBT were reviewed. Patients were irradiated on two linear accelerators, in one of two positions (standing, n = 77; reclined, n = 8) and 1830 in vivo TLD measurements were obtained for various locations on 76 patients.

RESULTS

The TLD results showed that the two TSEBT techniques were dosimetrically heterogeneous. At several sites, the dose administered correlated with height, weight, and gender. After the first TLD measurement, fourteen patients (18%) required MU modification, with a mean 10% reduction (range, -25 to +35). Individual TLD results allowed us to customize the boost treatment for each patient. For patients who were evaluated in the standing position, the most common underdosed sites were the axilla, perineum/perianal folds, and soles (each receiving 69%, 20%, and 34% of the prescribed dose, respectively). For patients evaluated in a reclining position, surface dose distribution was more heterogeneous. The sites underdosed most commonly were the axilla and perineum/perianal folds (receiving less than one third of prescribed dose). Significant variables were detected with model building.

CONCLUSION

TLD measurements were integral to quality assurance for TSEBT. Dose distribution at several anatomical sites correlated significantly with gender, height, and weight of the treated individual and might be predicted.

Для цитирования: Elsayad Khaled, Moustakis Christos, Simonsen Manuela, Bäcker Dagmar, Haverkamp Uwe, Eich Hans Theodor. In-vivo dosimetric analysis in total skin electron beam therapy.. Physics and imaging in radiation oncology. 2018. DOI: 10.1016/j.phro.2018.05.002.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.