Aortic annulus diameter measurement: what is the best modality?
Аннотация
PURPOSE: The purpose of this study was to clarify the most suitable method to measure the aortic annulus diameter.
PATIENTS AND
METHODS
Fifty-five patients, who had undergone aortic valve replacement at Miyazaki University Hospital between April 2008 and May 2011, were included in this study. The maximum diameter of the sizing tool that could be inserted into the left ventricle through the annulus had been predicted, based on the diameter measured by each modality. Agreement with surgery and each imaging modality, namely transthoracic echocardiography, multidetector computed tomography and contrast angiography, were compared using Bland-Altman analysis.
RESULTS
The predicted aortic annulus diameter, based on the diameter measured by transthoracic echocardiography, multidetector computed tomography and contrast angiography, was 20.3 ± 2.50, 23.9 ± 3.19, and 23.5 ± 3.55 mm, respectively, whereas, the diameter measured at surgery was 23.7 ± 2.99 mm. Predicted aortic annulus diameter measured by multidetector computed tomography best agreed with that measured at surgery.
CONCLUSION
We conclude that the aortic annulus diameter, measured by multidetector computed tomography, is the best modality to measure the aortic annulus diameter.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.