3D color-Doppler echocardiography and chronic aortic regurgitation: a novel approach for severity assessment.

Perez de Isla Leopoldo, Zamorano Jose, Fernandez-Golfin Covadonga, Ciocarelli Sara, Corros Cecilia, Sanchez Tibisai, Ferreirós Joaquín, Marcos-Alberca Pedro, Almeria Carlos, Rodrigo Jose Luis, Macaya Carlos

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

3D echocardiography provides a complete evaluation of the aortic valve and adjacent structures and it improves the assessment of this cardiac region. Three-dimensional color-Doppler echocardiography (3DCDE) evaluation might improve the measurements of the functional regurgitant orifice in patients with Chronic Aortic Regurgitation (CAR).

OBJECTIVES

Our aim was to compare the accuracy of current echo-Doppler methods and 3DCDE for the assessment of CAR severity. The reference method used in this work was the CAR severity determined by means of cardiac magnetic resonance (CMR)

METHODS

Thirty-two consecutive patients with an established diagnosis of CAR recruited in our institution comprised our study group. CAR severity was determined by conventional Echo-Doppler methods and by 3DCDE and their results were compared with those obtained by means of CMR.

RESULTS

Mean age was 63.0 ± 13.5 years. Twenty-two patients (68.8%) were men. Compared with the traditional echo-Doppler methods, 3DCDE evaluation had the best linear association with CMR results (3D vena contracta cross sectional area method: r = 0.88; r square = 0.77; p < 0.001. 3D vena contracta cross sectional area/left ventricular outflow tract cross sectional area method: r = 0.87; r square = 0.75; p < 0.001). The ROC analysis showed an excellent area under curve for detection of severe CAR (3D vena contracta cross sectional area method = 0.97; 3D vena contracta cross sectional area/left ventricular outflow tract cross sectional area method = 0.98). Inter- and intra-observer variability for the 3DCDE evaluation was good (ICC = 0.89 and ICC = 0.91 for inter and intra observer variability respectively).

CONCLUSIONS

3DCDE is an accurate and highly reproducible diagnostic tool for estimating CAR severity. Compared with the traditional echo-Doppler methods, 3DCDE has the best agreement with the CMR determined CAR severity. Thus, 3DCDE is a diagnostic method that may improve the therapeutic management of patients with CAR.

Для цитирования: Perez de Isla Leopoldo, Zamorano Jose, Fernandez-Golfin Covadonga, Ciocarelli Sara, Corros Cecilia, Sanchez Tibisai, Ferreirós Joaquín, Marcos-Alberca Pedro, Almeria Carlos, Rodrigo Jose Luis, Macaya Carlos. 3D color-Doppler echocardiography and chronic aortic regurgitation: a novel approach for severity assessment.. International journal of cardiology. 2011. DOI: 10.1016/j.ijcard.2011.11.094.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.