MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ejcts.2011.02.073

Медицинская библиотека · European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2011

In vitro investigation of aortic valve annuloplasty using prosthetic ring devices.

Scharfschwerdt Michael, Pawlik Mareike, Sievers Hans-Hinrich, Charitos Efstratios I

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The remodeling of the dilatated valve annulus with a prosthetic ring for the repair of valve insufficiency is a well-established concept in mitral valve surgery, and may also be suitable for aortic valve reconstruction. In this study, two models of prosthetic aortic annuloplasty devices were investigated.

METHODS

Fresh porcine aortic roots (n=16) were investigated in a pulsatile flow simulator after patch dilatation of the annulus and subsequent reconstruction using both an external and an internal prosthetic ring. For each configuration, leakage was determined by ultrasonic flow measurements and leaflet co-aptation by transesophageal echocardiography. In addition, valves' motions were recorded by high-speed video.

RESULTS

By the use of the prosthetic annuloplasty rings, leakage volumes decreased significantly compared with the dilatated root, more pronounced with the intra-annular ring. Similarly, the co-aptation height of the leaflets increased. Pressure gradients were not significantly influenced by the ring application, but leaflet motion patterns changed from the usual trapezoid to a more rectangular opening characteristic, visible at both echocardiographic and high-speed video analysis.

CONCLUSIONS

The reconstruction of a dilatated aortic valve annulus using external and internal ring devices is feasible and effective for reduction of regurgitation at which the internal ring provides a greater potential to decrease valve insufficiency.

Для цитирования: Scharfschwerdt Michael, Pawlik Mareike, Sievers Hans-Hinrich, Charitos Efstratios I. In vitro investigation of aortic valve annuloplasty using prosthetic ring devices.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2011. DOI: 10.1016/j.ejcts.2011.02.073.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.