DOI 10.1016/j.athoracsur.2010.12.050
New technique for aortic valve functional annulus reshaping using a handmade prosthetic ring.
Аннотация
BACKGROUND
Despite a wide development in aortic leaflets repair techniques, aortic valve annuloplastic procedures are still poorly investigated. We present our aortic valve annuloplastic system consisting of a handmade prosthetic ring with 2 components for reshaping the aortic annulus and sinotubular junction (STJ) and illustrates our surgical approach and clinical results.
METHODS
Since February 2003, 45 patients with aortic valve regurgitation underwent aortic annuloplasty using the new ring. Mean patient age was 58 ± 16 years (range, 46 to 76 years). The ring has 2 components: a circular ring used to undersize the circumference of the aortoventricular junction and 3-crown-like shape ring used for STJ remodeling. The circular ring was sutured into the left ventricular outflow tract in the aortic subvalvular position and the STJ ring was sutured from the outside of the aortic root just at the level of the STJ. The 3 vertical bands of the STJ ring were fixed to the underlying circular ring to stabilize the continuity between the STJ and nadir of the aortic valve.
RESULTS
No in-hospital death occurred. Two patients had residual trivial aortic valve regurgitation postoperatively, and 3 patients required treatment for residual aortic regurgitation (more than mild). Mean length of the coaptation surface was 10 ± 2 mm. The mean clinical follow-up (100% complete) was 22 ± 16 months. All patients were free from cardiac and valve-related events, and no complications due to ring implantation occurred.
CONCLUSIONS
This ring reshapes the functional aortic annulus and stabilizes all components of the native aortic valve stent to improve long-term results of valve repair.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.