BACKGROUND

Aortic valve integrity is crucial for optimal left ventricular assist device (LVAD) support. Pre-existing native aortic insufficiency, aortic valve incompetence acquired during support, as well as previously placed prosthetic aortic valves present unique problems for these patients.

METHODS

We reviewed and analyzed data for 28 patients who underwent left ventricular outflow tract closure associated with HeartMate I (n =12) and HeartMate II (n = 16) LVAD insertion or exchange. Indications for valve closure, surgical technique, LVAD function, survival rates and complications were retrospectively analyzed. Survival rates were compared with those of HeartMate LVAD patients (n = 104) who did not undergo aortic valve closure.

RESULTS

Indications for closure included native aortic valve insufficiency (10 patients), aortic valve deterioration after prolonged LVAD support (8 patients) and previously placed mechanical (9 patients) or bioprosthetic aortic prostheses (1 patient). There were 2 operative and 5 late deaths (mean 227 days post-operatively). Of the deaths, none were due to aortic valve closure. Actuarial survival was 78% at 1 year and 53% at 3 years, which was statistically better than for our patients with an intact aortic outflow (61% at 1 year, 45% at 3 years; p < 0.05). Five patients had transplants, 1 patient was successfully bridged to recovery, and 15 patients remain on LVAD support. No patient with outflow closure developed regurgitation, embolization or compromised LVAD support.

CONCLUSION

Outflow tract closure in LVAD-supported patients is safe, often necessary and well tolerated.

Для цитирования: Adamson Robert M, Dembitsky Walter P, Baradarian Sam, Chammas Joseph, May-Newman Karen, Chillcott Suzanne, Stahovich Marcia, McCalmont Vicki, Ortiz Kristi, Hoagland Peter, Jaski Brian. Aortic valve closure associated with HeartMate left ventricular device support: technical considerations and long-term results.. The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation. 2011. DOI: 10.1016/j.healun.2010.11.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.