Inappropriately long post-ventricular atrial blanking leading to 'unnecessary' defibrillator discharge.

Bera Debabrata, Kar Ayan, Das Arya Kamal, Majumder Suchit

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A 55-year-old lady with non-ischemic cardiomyopathy (NICM) was referred for multiple implantable cardioverter defibrillator (ICD) shocks. Stored electrograms (EGM) revealed atrial flutter (AFL) with A > V. Morphology match was good and RR-intervals were irregular. Despite all these, the dual-chamber-ICD (Abbott medical) classified this as ventricular tachycardia (VT-2) via V > A algorithm where it did not analyze morphology/stability and delivered therapy. Anti-tachycardia-pacing (ATP) was delivered which induced a true VT (rate in VF-zone) and immediate shock was delivered. It was hence appropriate but an 'unnecessary' shock. The offender was found to be an inappropriately programmed long post-ventricular atrial-blanking (PVAB) of 200 ms which led to undersensing of several atrial electrograms, falsely making V > A during a clear AFL.

Для цитирования: Bera Debabrata, Kar Ayan, Das Arya Kamal, Majumder Suchit. Inappropriately long post-ventricular atrial blanking leading to 'unnecessary' defibrillator discharge.. Pacing and clinical electrophysiology : PACE. 2022. DOI: 10.1111/pace.14506.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.