Ultrasound as the imaging method of choice for monitoring the intragastric balloon in obese patients: normal findings, pitfalls and diagnosis of complications.

Francica Giampiero, Giardiello Cristiano, Iodice Giuseppe, Cristiano Stefano, Scarano Ferdinando, Delle Cave Massimo, Sarrantonio Gennaro, Troiano Ersilia, Cerbone Maria Rosaria

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The authors studied ultrasound (U/S) monitoring of the BioEnterics intragastric balloon (BIB) for treatment of moderate obesity by describing normal U/S aspects, diagnostic pitfalls and assessment of complications.

METHODS

Over the last 3 years, 151 BIB systems have been endoscopically placed and filled with 500-700 ml of saline plus 10 ml of methylene blue in 131 patients (mean age 38.4 years, range 18-72); males/females 46/85; mean BMI 43.8 kg/m(2)). In all patients abdominal U/S was carried out before and 12 days after endoscopy and thereafter at scheduled follow-up examinations, and/or whenever the subject complained of epigastric burning/ache, vomiting, and lack of sensation of a BIB with or without staining of urine and/or feces.

RESULTS

The BIB appeared as a round anechoic structure, with a thick hyperechoic wall and a hyperechoic band-like valve inside. Complications occurred in 18 patients: in 8, staining of urine and/or feces prompted immediate endoscopic removal. In 10 patients, U/S findings were: 1) decreased volume, loss of the spherical shape with the valve still visible (5 cases); 2) multiple hyperechoic streaks regularly spaced due to a completely collapsed BIB, not modified by decubitus changes (3 cases); 3) migration through the intestine (2 cases): in one patient the BIB was passed in the stools after 4 days, whereas in the other case large bowel obstruction required laparoscopic surgery.

CONCLUSION

U/S affords a quick, safe and accurate method for assessing both BIB status after endoscopic deployment and the most frequently-occurring complications (partial/complete deflation, migration through the pylorus with intestinal obstruction) in obese patients.

Для цитирования: Francica Giampiero, Giardiello Cristiano, Iodice Giuseppe, Cristiano Stefano, Scarano Ferdinando, Delle Cave Massimo, Sarrantonio Gennaro, Troiano Ersilia, Cerbone Maria Rosaria. Ultrasound as the imaging method of choice for monitoring the intragastric balloon in obese patients: normal findings, pitfalls and diagnosis of complications.. Obesity surgery. 2004. DOI: 10.1381/0960892041590791.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.