DOI 10.1111/jce.15379

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2022

Comparison of ablation index versus time-guided radiofrequency energy dosing using normal and half-normal saline irrigation in a porcine left ventricular model.

Larsen Timothy, Du-Fay-de-Lavallaz Jeanne M, Winterfield Jeffrey R, Ravi Venkatesh, Rhodes Paul, Wasserlauf Jeremiah, Trohman Richard G, Sharma Parikshit S, Huang Henry D

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Ablation Index (AI) is a multiparametric quality marker to assess the durability of radiofrequency (RF) lesions. The comparative effectiveness and safety of AI versus time-based energy dosing for ablation of ventricular arrhythmias are unknown.

OBJECTIVE

We compared AI and time-based RF dosing strategies in the left ventricles (LVs) of freshly harvested porcine hearts.

METHODS

Ablation was performed in vitro with an open-irrigated ablation catheter (Thermocool ST/SF), 40 W, contact force 10-15 g. Tissue samples were stained in triphenyltetrazolium chloride for the measurement of lesion dimensions.

RESULTS

A total of 560 lesions were performed (AI-group: [n = 360]; time-group: [n = 200]). Using normal saline (NS) (n = 280), growth in lesion depth slowed after 30 s and AI > 550 in comparison to width, volume, and magnitude of impedance drops which continued to increase with longer RF duration. Risk of steam pop (SP) was higher for RF > 30 s (RF < 30 s:1 SP [2.5%] vs. RF > 30 s: 15 SP [25%]; p = .002) or AI targets >550 (AI: 350-550: 2 SP [2%] vs. AI 600-750: 15 SP [19%]; p = .001). Using half-normal saline (HNS) (n = 280), lesion dimension and impedance drops were larger and growth in lesion depth slowed earlier (AI: 500). Risk of SPs was higher above AI 550 (AI: 350-550: 7 [7%] SPs vs. AI 600-750: 28 [35%] SPs; p < .00001). While codependent variables, correlation between AI and time was modest-to-strong but decreased with longer RF duration.

CONCLUSION

In this ex vivo study, AI was a better predictor of lesion dimensions than ablation time and magnitude of impedance drop in the LV using NS and HNS irrigation. AI targets above 550 led to a higher risk of SPs. Future trials are required to verify these findings.

Для цитирования: Larsen Timothy, Du-Fay-de-Lavallaz Jeanne M, Winterfield Jeffrey R, Ravi Venkatesh, Rhodes Paul, Wasserlauf Jeremiah, Trohman Richard G, Sharma Parikshit S, Huang Henry D. Comparison of ablation index versus time-guided radiofrequency energy dosing using normal and half-normal saline irrigation in a porcine left ventricular model.. Journal of cardiovascular electrophysiology. 2022. DOI: 10.1111/jce.15379.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.