DOI 10.1177/19322968261436412

Медицинская библиотека · Journal of diabetes science and technology · 2026

First in Human Feasibility Study of an Insulin Patch Pump Combined With CGM-Insulin Delivery Cannula.

Cunningham Hannah, Jones Hanna, Flint Steve, Netzer Emma, Dutt-Ballerstadt Ralph, Eckenberg Peter, Obeyesekere Varuni, Brown Katrin Ellen, Sims Catriona, Reid Solomon, Wu Huan-Ping, Seidl Thomas, Both Marcel, Auderset Adrian, Eymann Daniel, Lodico Tamara, Schoemaker Michael, Kong Yee Wen, O'Neal David Norman

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Combining a patch pump (PP) with a single-insertion glucose-sensing cannula (continuous glucose monitor with an insulin set [CGM-IS]) may reduce user burden compared with tethered insulin pumps or multiple daily injections. We aimed to determine feasibility of a combined device.

METHODS

This feasibility study evaluated a PP with a CGM-IS (PP-CGM-IS) with manual bolus insulin dosing in T1D adults with insulin pump and CGM experience. The pilot phase involved 3 participants undergoing a mixed meal test (MMT) for 12 hours. In the second phase, 15 participants were studied for 72 hours, including day 1 and day 3 MMTs and one supervised free-living day at a hotel. Primary outcome was accuracy of the CGM device, assessed by mean absolute relative difference (MARD). Secondary outcomes included number of insulin delivery failures and device survival duration.

RESULTS

Twenty-six devices were inserted in 18 participants. Mean age was 51 years, 14 were female. Nine device (sensor) failures occurred immediately post-insertion requiring re-insertion. Mean absolute relative difference against Yellow Springs Instruments (YSI) in 17 devices was 11.6%. Consensus error grid analysis against YSI analyser showed 83.16% of readings fell within zone A and 100% within zones A and B. Mean glucose, time in range, and total daily insulin dose on day 2 of the main phase compared with baseline did not suggest compromise of insulin delivery. There were no serious adverse events.

CONCLUSION

This first-in-human study confirms feasibility of integrating CGM-IS with PP. A PP-CGM-IS system warrants further research, including improved insertion process, greater sensor accuracy and ultimately real-time algorithm implementation.

Для цитирования: Cunningham Hannah, Jones Hanna, Flint Steve, Netzer Emma, Dutt-Ballerstadt Ralph, Eckenberg Peter, Obeyesekere Varuni, Brown Katrin Ellen, Sims Catriona, Reid Solomon, Wu Huan-Ping, Seidl Thomas, Both Marcel, Auderset Adrian, Eymann Daniel, Lodico Tamara, Schoemaker Michael, Kong Yee Wen, O'Neal David Norman. First in Human Feasibility Study of an Insulin Patch Pump Combined With CGM-Insulin Delivery Cannula.. Journal of diabetes science and technology. 2026. DOI: 10.1177/19322968261436412.