Implanted Electronic Device Management in a Pencil-Beam Scanning Proton-Therapy Clinic.

Pepin Mark D, Ball Brooke E, Stuart-Mullen Lynette G, Schroeder Sarah J, Klein Kristi A, Zibrowski Megan R, Brost Eric E, Brinkmann Debra H, Remmes Nicholas B, Garces Yolanda I, Mulpuru Siva K, Hallemeier Christopher L

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: This study aims to present single-institution workflows and experience for treating patients with implanted electronic devices (IEDs) at a high-volume proton pencil-beam scanning clinic.

METHODS AND MATERIALS: A multidisciplinary group was formed when opening the proton clinic to assess the risk of IEDs, primarily from neutron damage, and the potential for patient harm under device failure. A policy was developed for management of cardiac (CIEDs) and medical IEDs (MIEDs), including endocrinology devices, neurostimulators, auditory-assist devices, and infusion pumps. Patients treated with IEDs were assessed for trends and failures.

RESULTS

A workflow was developed to identify devices and communicate potential action steps at 4 stages of treatment planning and delivery. Guidelines on-treatment beam selection and image guidance were shared between all device types. Patients with CIED received heart rate monitoring during each treatment and regular device interrogations, while guidance for MIEDs varied by function and risk of harm. A total of 294 proton radiation therapy courses were delivered to patients with IEDs. Of 144 courses delivered to patients with CIED devices, 5 had recoverable resets with no major failures.

CONCLUSIONS

Risk-based guidelines were developed and refined to safely treat patients' CIEDs and MIEDs.

Для цитирования: Pepin Mark D, Ball Brooke E, Stuart-Mullen Lynette G, Schroeder Sarah J, Klein Kristi A, Zibrowski Megan R, Brost Eric E, Brinkmann Debra H, Remmes Nicholas B, Garces Yolanda I, Mulpuru Siva K, Hallemeier Christopher L. Implanted Electronic Device Management in a Pencil-Beam Scanning Proton-Therapy Clinic.. Practical radiation oncology. 2026. DOI: 10.1016/j.prro.2026.03.004.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.