DOI 10.1148/radiology.211.3.r99jn42775
Carotid arterial stenosis: evaluation at CT angiography with the volume-rendering technique.
Аннотация
PURPOSE: To determine whether computed tomographic (CT) angiography with the volume-rendering technique (VRT) can be used to accurately quantify carotid arterial stenosis and to identify occlusions.
MATERIALS AND
METHODS
Spiral CT was performed in 23 patients who were referred for carotid stenosis evaluation. VRT images and shaded-surface display (SSD) images of 46 carotid arterial bifurcations were compared with findings from digital subtraction angiography (DSA).
RESULTS
Agreement on stenosis category between VRT CT angiography and DSA was found in 39 (85%) of the 46 carotid arteries studied. VRT CT angiography was 92% (49 of 53) sensitive and 96% (82 of 85) specific for the detection of grade 2-3 stenoses (> or = 70% stenosis). Agreement on stenosis category between SSD CT angiography and DSA was found in 38 (83%) of the 46 carotid arteries studied. SSD CT angiography was 91% (48 of 53) sensitive and 93% (79 of 85) specific for the detection of grade 2-3 stenoses. Calcified stenoses were correctly graded at VRT CT angiography in 10 of the 10 cases with heavy mural calcified plaques, while eight of the 10 stenoses were accurately quantified at SSD CT angiography.
CONCLUSION
These results indicate that VRT CT angiography is as accurate as SSD CT angiography in the evaluation of carotid arterial bifurcations.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.