Detection of individual responses to clopidogrel: Validation of a novel, rapid analysis using thrombelastography 6s.
Аннотация
INTRODUCTION
There is potential value in testing individual response to P2Y12 inhibitors to predict ischemic and bleeding risk in patients undergoing percutaneous coronary intervention. The aims of this study were: (1) to validate the ability of a novel point of care (POC) assay, thrombelastography (TEG) 6s, to detect changes in adenosine diphosphate (ADP)-induced whole blood clotting in volunteers and patients given clopidogrel using TEG 5000 as a reference and (2) to compare a novel, rapid parameter, area under the curve at 15 minutes (AUC15), with the traditional maximum clot amplitude (MA) in TEG 6s.
METHODS
A total of 25 participants were included in whom ADP-induced clotting was measured at 4 time points: (1) 12 healthy volunteers given 600 mg of clopidogrel; (2) 12 patients with ACS given 600 mg of clopidogrel; (3) 1 healthy volunteer given 600 mg of clopidogrel on 5 separate occasions. All samples were tested using conventional TEG 5000 and the new POC TEG 6S, and a new parameter called AUC15 was compared with MA in TEG 6s.
RESULTS
(1) TEG 5000 and TEG 6s both detected changes in ADP-induced platelet activation. Bland-Altman analysis demonstrated a good level of agreement between them. (2) For TEG 6S, correlation between MA and the novel AUC15 was strong for both thrombin and ADP channels (R2 = 0.867, R = .936, P < .001), and the AUC15 result was available on average 13.3 minutes earlier.
CONCLUSIONS
Thrombelastography 6s is a rapid, easy to use and accurate test of ADP-induced clotting using TEG 5000 as a reference. A novel parameter, AUC15, is a viable, time-saving option for this test and has potential value in personalized P2Y12 inhibitor therapy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.