Early structural valve deterioration and reoperation associated with the mitroflow aortic valve.

Axtell Andrea L, Chang David C, Melnitchouk Serguei, Jassar Arminder S, Tolis George, Villavicencio Mauricio A, Sundt Thoralf M, D'Alessandro David A

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Structural valve deterioration (SVD) is a known limitation of bioprosthetic valves. Recent reports have suggested a concerning rate of early SVD in patients receiving a Mitroflow aortic bioprosthesis. We therefore compared the incidence of SVD and SVD requiring reoperation among patients receiving a Mitroflow versus a common contemporary bioprosthesis.

METHODS

A retrospective cohort analysis was performed on 592 patients receiving a Mitroflow aortic bioprosthesis at our institution between 2010 and 2014. Patients were matched 1:1 using a coarsened exact matching algorithm with patients receiving a Carpentier-Edwards Magna Ease aortic bioprosthesis (Edwards Lifesciences, Irvine, CA) during the same period. The incidence of SVD (defined as a mean transprosthetic gradient ≥30 mmHg or moderate to severe intraprosthetic regurgitation), reoperation for SVD, and cumulative survival were compared between prosthesis types.

RESULTS

The cumulative incidence of SVD at 5 years for all patients receiving a Mitroflow aortic bioprosthesis was 16% (13-21%) and 5% underwent reoperation for SVD. Implantation of a Mitroflow valve was associated with an increased risk of SVD compared to the comparator valve (hazard ratio [HR] 2.59 [1.69-3.98], P < 0.01). Older age had a protective effect against SVD (HR 0.95 [0.93-0.96], P < 0.01). Patients who received a Mitroflow valve had reduced long-term survival compared to those who received a comparator valve (P = 0.03).

CONCLUSION

The Mitroflow aortic bioprosthesis is associated with increased rates of early SVD and reoperation for valvular dysfunction as well as reduced survival compared to a contemporary valve. Enhanced clinical and echocardiographic follow-up is advisable after Mitroflow implantation.

Для цитирования: Axtell Andrea L, Chang David C, Melnitchouk Serguei, Jassar Arminder S, Tolis George, Villavicencio Mauricio A, Sundt Thoralf M, D'Alessandro David A. Early structural valve deterioration and reoperation associated with the mitroflow aortic valve.. Journal of cardiac surgery. 2018. DOI: 10.1111/jocs.13953.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.