MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2018.05.023

Медицинская библиотека · JACC. Cardiovascular interventions · 2018

Percutaneous Left Atrial Appendage Closure With the Ultraseal Device: Insights From the Initial Multicenter Experience.

Asmarats Lluis, Masson Jean-Bernard, Pagnotta Paolo A, Cook Stéphane, Foresti Mike, Ibrahim Réda, Sukiennik Adam, Sabiniewicz Robert, Maffeo Diego, Carballo Julio, Cruz-González Ignacio, Grasso Carmelo, Pisano Francesco, Senatore Gaetano, Tarantini Giuseppe, Kasongo Adolphe, Chiarito Mauro, Puricel Serban, Messas Nathan, Moreno-Samos Jose Carlos, O'Hara Gilles, Rodés-Cabau Josep

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to evaluate the feasibility, safety, and efficacy of the Ultraseal device for left atrial appendage closure (LAAC) (Cardia, Eagan, Minnesota) in patients with nonvalvular atrial fibrillation at high bleeding risk.

BACKGROUND

The Ultraseal device is a novel bulb-and-sail designed LAAC device, with an articulating joint enabling conformability to heterogeneous angles and shapes of appendage anatomy.

METHODS

This was a multicenter study including consecutive patients undergoing LAAC with the Ultraseal device at 15 Canadian and European sites. Periprocedural and follow-up events were systematically collected, and transesophageal echocardiography at 45 to 180 days post-procedure was routinely performed in all centers but 3.

RESULTS

A total of 126 patients (mean age 75 ± 8 years; mean CHA2DS2-VASc score 5 ± 2; mean HAS-BLED score 4 ± 1) were included. The device was successfully implanted in 97% of patients. A major periprocedural adverse event occurred in 3 (2.4%) patients (clinically relevant pericardial effusion [n = 1], stroke [n = 1], device embolization [n = 1]). Ninety percent of patients were discharged on single or dual antiplatelet therapy. Follow-up transesophageal echocardiography was available in 89 (73%) patients, with no cases of large (>5 mm) residual leak and 5 (5.6%) cases of device-related thrombosis (all successfully treated with anticoagulation therapy). At a median follow-up of 6 (interquartile range: 3 to 10) months, the rates of stroke and transient ischemic attack were 0.8% and 0.8%, respectively, with no systemic emboli. None of the events occurred in patients with device-related thrombosis.

CONCLUSIONS

In this initial multicenter experience, LAAC with the Ultraseal device was associated with a high implant success rate and a very low incidence of periprocedural complications. There were no late device-related clinical events and promising efficacy results were observed regarding thromboembolic prevention at midterm follow-up. Larger studies are further warranted to confirm the long-term safety and efficacy of this novel device.

Для цитирования: Asmarats Lluis, Masson Jean-Bernard, Pagnotta Paolo A, Cook Stéphane, Foresti Mike, Ibrahim Réda, Sukiennik Adam, Sabiniewicz Robert, Maffeo Diego, Carballo Julio, Cruz-González Ignacio, Grasso Carmelo, Pisano Francesco, Senatore Gaetano, Tarantini Giuseppe, Kasongo Adolphe, Chiarito Mauro, Puricel Serban, Messas Nathan, Moreno-Samos Jose Carlos, O'Hara Gilles, Rodés-Cabau Josep. Percutaneous Left Atrial Appendage Closure With the Ultraseal Device: Insights From the Initial Multicenter Experience.. JACC. Cardiovascular interventions. 2018. DOI: 10.1016/j.jcin.2018.05.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.