Analysis of closed-cell intracranial stent characteristics using cone-beam computed tomography with contrast material.

Tsuruta Wataro, Matsumaru Yuji, Hamada Yusuke, Hayakawa Mikito, Kamiya Yuki

СтатьяПеревод пока не подготовлен

Аннотация

The intracranial nitinol stent named the Enterprise Vascular Reconstruction Device has poor radiographic visibility. The characteristics of closed-cell intracranial stents were investigated and the efficacy of intraoperative stent visualization examined with the 80 kV high-resolution XperCT protocol, which is a flat detector C-arm volume acquisition functionality system integrated with the angiography equipment. We treated 39 aneurysms with stent-assisted coil embolization. The aneurysms were located on the internal carotid artery in 24 cases, the anterior communicating artery (AcomA) in three, the basilar artery (BA) in 10, and the vertebral artery in two. Intraoperative 80 kV XperCT was performed in all cases after deposition of the stent. We evaluated the coverage of the aneurysm neck, incomplete stent apposition (ISA), and shift of vessels. Accurate stent visualization was achieved in 29 of the 39 cases without coil and delivery wire artifact. Coverage of the aneurysm neck succeeded in 28 cases; there was one case of BA top Y-configuration stenting in which the stent was dislocated into the aneurysm. ISA was detected in nine cases, including seven kinks and one flattening in the carotid siphon and one kink in the BA top. We detected linearization of vessels due to stent deployment in three AcomA cases and three BA top cases. We conclude that intraoperative 80 kV XperCT is an efficient modality for the evaluation of ISA. Stent kinking in the carotid siphon and linearization in distal vessels can be detected with this protocol.

Для цитирования: Tsuruta Wataro, Matsumaru Yuji, Hamada Yusuke, Hayakawa Mikito, Kamiya Yuki. Analysis of closed-cell intracranial stent characteristics using cone-beam computed tomography with contrast material.. Neurologia medico-chirurgica. 2013. DOI: 10.2176/nmc.53.403.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.