DOI 10.1016/s0140-6736(23)02189-x

Медицинская библиотека · Lancet (London, England) · 2023

Mortality in randomised controlled trials using paclitaxel-coated devices for femoropopliteal interventional procedures: an updated patient-level meta-analysis.

Parikh Sahil A, Schneider Peter A, Mullin Christopher M, Rogers Tyson, Gray William A

МетаанализСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Numerous randomised clinical trials and real-world studies have supported the safety of paclitaxel-coated devices for the treatment of femoropopliteal occlusive disease. However, a 2018 summary-level meta-analysis suggested an increased mortality risk for paclitaxel-coated devices compared with uncoated control devices. This study presents an updated analysis of deaths using the most complete and current data available from pivotal trials of paclitaxel-coated versus control devices.

METHODS

Ten trials comparing paclitaxel-coated versus control devices were included in a patient-level pooled analysis. Cox regression models were used to evaluate the effect of paclitaxel exposure on risk of death in both intention-to-treat (ITT; primary analysis) and three as-treated analysis sets accounting for treatment group crossover at the index procedure and over time. The effect of paclitaxel dose and baseline covariates were also evaluated.

FINDINGS: A total of 2666 participants were included with a median follow-up of 4·9 years. No significant increase in deaths was observed for patients treated with paclitaxel-coated devices. This was true in the ITT analysis (hazard ratio [HR] 1·14, 95% CI 0·93-1·40), the as-treated analysis (HR 1·13, 95% CI 0·92-1·39), and in two crossover analyses: 1·07 (0·87-1·31) when late crossovers were censored and 1·04 (0·84-1·28) when crossovers were analysed from the date of paclitaxel exposure. There was no significant effect of paclitaxel dose on mortality risk.

INTERPRETATION: This meta-analysis found no association between paclitaxel-coated device exposure and risk of death, providing reassurance to patients, physicians, and regulators on the safety of paclitaxel-coated devices.

FUNDING: Becton Dickinson, Boston Scientific, Cook, Medtronic, Philips, Surmodics, and TriReme Medical.

Для цитирования: Parikh Sahil A, Schneider Peter A, Mullin Christopher M, Rogers Tyson, Gray William A. Mortality in randomised controlled trials using paclitaxel-coated devices for femoropopliteal interventional procedures: an updated patient-level meta-analysis.. Lancet (London, England). 2023. DOI: 10.1016/s0140-6736(23)02189-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.