DOI 10.1016/j.jacc.2011.07.015
Microembolization during carotid artery stenting in patients with high-risk, lipid-rich plaque. A randomized trial of proximal versus distal cerebral protection.
Аннотация
OBJECTIVES
The goal of this study was to compare the rate of cerebral microembolization during carotid artery stenting (CAS) with proximal versus distal cerebral protection in patients with high-risk, lipid-rich plaque.
BACKGROUND
Cerebral protection with filters partially reduces the cerebral embolization rate during CAS. Proximal protection has been introduced to further decrease embolization risk.
METHODS
Fifty-three consecutive patients with carotid artery stenosis and lipid-rich plaque were randomized to undergo CAS with proximal protection (MO.MA system, n = 26) or distal protection with a filter (FilterWire EZ, n = 27). Microembolic signals (MES) were assessed by using transcranial Doppler during: 1) lesion wiring; 2) pre-dilation; 3) stent crossing; 4) stent deployment; 5) stent dilation; and 6) device retrieval/deflation. Diffusion-weighted magnetic resonance imaging was conducted before CAS, after 48 h, and after 30 days.
RESULTS
Patients in the MO.MA group had higher percentage diameter stenosis (89 ± 6% vs. 86 ± 5%, p = 0.027) and rate of ulcerated plaque (35% vs. 7.4%; p = 0.019). Compared with use of the FilterWire EZ, MO.MA significantly reduced mean MES counts (p < 0.0001) during lesion crossing (mean 18 [interquartile range (IQR): 11 to 30] vs. 2 [IQR: 0 to 4]), stent crossing (23 [IQR: 11 to 34] vs. 0 [IQR: 0 to 1]), stent deployment (30 [IQR: 9 to 35] vs. 0 [IQR: 0 to 1]), stent dilation (16 [IQR: 8 to 30] vs. 0 [IQR: 0 to 1]), and total MES (93 [IQR: 59 to 136] vs. 16 [IQR: 7 to 36]). The number of patients with MES was higher with the FilterWire EZ versus MO.MA in phases 3 to 5 (100% vs. 27%; p < 0.0001). By multivariate analysis, the type of brain protection was the only independent predictor of total MES number. No significant difference was found in the number of patients with new post-CAS embolic lesion in the MO.MA group (2 of 14, 14%) as compared with the FilterWire EZ group (9 of 21, 42.8%).
CONCLUSIONS
In patients with high-risk, lipid-rich plaque undergoing CAS, MO.MA led to significantly lower microembolization as assessed by using MES counts.