MedElevator Войти в исследования
← Поиск публикаций

DOI 10.6009/jjrt.25-1509

Медицинская библиотека · Nihon Hoshasen Gijutsu Gakkai zasshi · 2025

[Evaluation of the Effectiveness of Bladder Ultrasound Imaging Devices in Radiation Therapy].

Kawasaki Yoshiyuki, Yoshida Tatsuya, Hayakawa Tomoki, Kawadai Toshiyuki, Shibasaki Takako

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: In radiation therapy for prostate cancer, bladder ultrasound diagnostic devices are utilized. However, the performance of all devices has not been fully evaluated. This study aims to assess the effectiveness of three devices-Lilium IP200 (IP200; Lilium Otsuka, Kanagawa), CubeScan BioCon-900S (900S; JCT, Hiroshima), and BVI 6100 (BVI 6100; Verathon, Bothell, WA, USA)-as tools for measuring bladder volume.

METHODS

The accuracy of each measurement device was verified by comparing the measurements taken with each device during treatment planning computed tomography (CT) with the reference values obtained from the bladder region of interest (ROI) created from the images taken immediately after the CT.

RESULTS

The differences from the reference values were 15.0±47.4 ml for the IP200, -14.7±42.8 ml for the 900S, and -4.2±32.2 ml for the BVI 6100. The Spearman's rank correlation coefficients were 0.78 for the IP200, 0.87 for the 900S, and 0.91 for the BVI 6100, with no significant differences observed between the measurement devices (p<0.05), indicating a strong correlation with the reference values.

CONCLUSION

The IP200, 900S, and BVI 6100 are capable of accurately and conveniently assessing bladder volume in radiation therapy.

Для цитирования: Kawasaki Yoshiyuki, Yoshida Tatsuya, Hayakawa Tomoki, Kawadai Toshiyuki, Shibasaki Takako. [Evaluation of the Effectiveness of Bladder Ultrasound Imaging Devices in Radiation Therapy].. Nihon Hoshasen Gijutsu Gakkai zasshi. 2025. DOI: 10.6009/jjrt.25-1509.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.