Technical Note: Significant Positive Bias in Lower Concentrations but Negative Bias in Higher Concentrations in Testosterone Measurement Using Beckman Access Immunoassay Compared to a Liquid Chromatography-Tandem Mass Spectrometry Reference Method.

Dasgupta Amitava, Woodard Kelsey, Ramoo Bheemraj, Frazee Clinton, Garg Uttam

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Testosterone is the principal male sex hormone and is secreted primarily by the testes. In most clinical laboratories testosterone is routinely measured for diagnosis of male hypogonadism or androgen excess in females using FDA approved immunoassays. We compared testosterone values measured by Beckman access immunoassay with those measured by a reference LC-MS/MS method.

METHODS

Testosterone was measured in 36 patients using left over serum or plasma specimens by both Beckman immunoassay on the DXI 800 analyzer and a reference LC-MS/MS method.

RESULTS

We observed overall significant negative bias of approximately 31.9 % when testosterone values obtained by the reference LC-MS/MS method were plotted in the x-axis and the corresponding testosterone values using the immunoassay in the y-axis, as regression equation was y=0.6887x+38.81 (n=36). The corresponding Deming regression was y=0.6639x+34.7163. However, in eight specimens with low testosterone concentrations, immunoassays significantly overestimated testosterone concentrations.

CONCLUSIONS

Immunoassays may underestimate the true testosterone concentration in males but overestimate in females with low testosterone concentration. Therefore, for diagnosis of hypogonadism in males and androgen excess in females, testosterone values obtained by Beckman Access immunoassay on the DXI 800 analyzer should be interpreted with caution.

Для цитирования: Dasgupta Amitava, Woodard Kelsey, Ramoo Bheemraj, Frazee Clinton, Garg Uttam. Technical Note: Significant Positive Bias in Lower Concentrations but Negative Bias in Higher Concentrations in Testosterone Measurement Using Beckman Access Immunoassay Compared to a Liquid Chromatography-Tandem Mass Spectrometry Reference Method.. Annals of clinical and laboratory science. 2024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.