Portal vein velocities measured by ultrasound: usefulness for evaluating shunt functioning following TIPS placement and TIPS revision.
Аннотация
BACKGROUND
To assess the usefulness of following portal vein velocities by Doppler ultrasound in evaluating shunt functioning after transhepatic intrajugular portosystemic shunt (TIPS) placement and revision.
METHODS
We retrospectively analyzed 39 patients who underwent a TIPS procedure in the preceding 4 years. Portal vein (PV) velocities were measured by Doppler ultrasound before and after TIPS insertion and revision and were correlated with portosystemic gradients (PSG) measured at angiography.
RESULTS
Mean PV velocities increased from 18 +/- 6 cm/s before TIPS placement to 50 +/- 21 cm/s (p < 0.001) after TIPS placement, with corresponding decrease of mean PSG from 20 +/- 6 to 8 +/- 3 mmHg (p < 0.001). Mean PV velocities significantly increased from 24 +/- 6 to 43 +/- 14 cm/s after TIPS revision (p < 0.02), with decrease of PSG from 17 +/- 6 to 9 +/- 5 mmHg (p < 0.05). A significant correlation was found between all PV velocities and their corresponding PSG (Spearman r < 0.61, p < 0.001).
CONCLUSION
PV velocities significantly correlate with changes in PSG. Following portal vein velocities by ultrasound is useful for early evaluation of shunt functioning following TIPS.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.