DOI 10.1097/iae.0000000000001735
HANDHELD SPECTRAL DOMAIN OPTICAL COHERENCE TOMOGRAPHY IMAGING THROUGH THE UNDILATED PUPIL IN INFANTS BORN PRETERM OR WITH HYPOXIC INJURY OR HYDROCEPHALUS.
Аннотация
PURPOSE: The authors investigated feasibility of undilated handheld spectral domain optical coherence tomography (SDOCT) retinal imaging in preterm infants and children with neurologic abnormalities.
METHODS
Under an institutional review board-approved protocol, the authors attempted handheld SDOCT imaging of the retina, choroid, and optic nerve in infants and young children without pupil dilation. Scans were analyzed for quality and successful capture of foveal, optic nerve, and retinal structural parameters and abnormalities.
RESULTS
The authors obtained images through an undilated pupil of 11 infants/children over 28 eye imaging sessions, 27 at the bedside without sedation, and one under anesthesia. Infants had retinopathy of prematurity (n = 8), hypoxic ischemic encephalopathy (n = 2), or obstructive hydrocephalus (n = 1 child). Pupil sizes ranged from 1.0 mm to 3.5 mm. The authors captured fovea and optic nerve scans in 25/28 eye imaging sessions, with scans of adequate quality to discern prespecified foveal and optic nerve morphology, and of the 25 sessions, the choroidal-scleral junction was visible in all but 6 sessions.
CONCLUSION
Undilated, handheld SDOCT imaging is a potential alternative method to evaluate the retina and optic nerve in patients with relative contraindication to pharmacological pupil dilation. This approach will enable the study of the eye-brain connection and ocular manifestations of neurologic diseases.