Neurogastroenterology and motility · 2021
Manometric parameters, when measured with the 3-dimensional high-definition anorectal manometry probe, poorly predict prolonged balloon expulsion time.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Woo Matthew, Pandey Armaan, Gill Harman, Li Dorothy, Buresi Michelle, Nasser Yasmin, Sawhney Summit, Andrews Christopher N
Аннотация
BACKGROUND: There are little data evaluating the performance of the 3-dimensional high-definition anorectal manometry (3D-HDAM) system in the diagnosis of dyssynergic defecation. Physical properties of the thicker, rigid, 3D-HDAM probe may have implications on the measurements of anorectal pressures.
AIM: Our aim was to compare 3D-HDAM to balloon expulsion test and magnetic resonance (MR) defecography.
METHODS: Consecutive constipated patients referred for anorectal function testing at the Calgary Gut Motility Centre (Calgary, Canada) between 2014 and 2019 were assessed. All patients underwent anorectal manometry with the 3D-HDAM probe, and a subset underwent BET or MR defecography. Anorectal manometric variables were compared between patients who had normal and abnormal BET.
RESULTS: Over the study period, 81 patients underwent both 3D-HDAM and BET for symptoms of constipation. 52 patients expelled the balloon within 3 minutes. Patients with abnormal BET had significantly lower rectoanal pressure differential (RAPD) (-61 vs. -31 mmHg for normal BET, p = 0.03) and defecation index (0.29 vs. 0.56, p = 0.03). On logistic regression analysis, RAPD (OR: 0.99, 95% CI: 0.97-0.99, p = 0.03) remained a negative predictor of abnormal BET. On ROC analysis, RAPD had an AUC of 0.65. There was good agreement between dyssynergic patterns on 3D-HDAM and defecographic evidence of dyssynergia (sensitivity 80%, specificity 90%, PLR 9, NLR 0.22, accuracy 85%).
CONCLUSIONS: Manometric parameters, when measured with the 3D-HDAM probe, poorly predict prolonged balloon expulsion time. RAPD remains the best predictor of prolonged balloon expulsion time. The 3D-HDAM probe may not be the ideal tool to diagnose functional defecatory disorders.
AIM: Our aim was to compare 3D-HDAM to balloon expulsion test and magnetic resonance (MR) defecography.
METHODS: Consecutive constipated patients referred for anorectal function testing at the Calgary Gut Motility Centre (Calgary, Canada) between 2014 and 2019 were assessed. All patients underwent anorectal manometry with the 3D-HDAM probe, and a subset underwent BET or MR defecography. Anorectal manometric variables were compared between patients who had normal and abnormal BET.
RESULTS: Over the study period, 81 patients underwent both 3D-HDAM and BET for symptoms of constipation. 52 patients expelled the balloon within 3 minutes. Patients with abnormal BET had significantly lower rectoanal pressure differential (RAPD) (-61 vs. -31 mmHg for normal BET, p = 0.03) and defecation index (0.29 vs. 0.56, p = 0.03). On logistic regression analysis, RAPD (OR: 0.99, 95% CI: 0.97-0.99, p = 0.03) remained a negative predictor of abnormal BET. On ROC analysis, RAPD had an AUC of 0.65. There was good agreement between dyssynergic patterns on 3D-HDAM and defecographic evidence of dyssynergia (sensitivity 80%, specificity 90%, PLR 9, NLR 0.22, accuracy 85%).
CONCLUSIONS: Manometric parameters, when measured with the 3D-HDAM probe, poorly predict prolonged balloon expulsion time. RAPD remains the best predictor of prolonged balloon expulsion time. The 3D-HDAM probe may not be the ideal tool to diagnose functional defecatory disorders.